ABSTRACT: Purpose
Total marrow (and lymphoid) irradiation (TMI-TMLI) is limited by the couch travel range of modern linacs, which forces the treatment delivery to be split into two plans with opposite orientations: a head-first supine upper-body plan, and a feet-first supine lower extremities plan. A specific field junction is thus needed to obtain adequate target coverage in the overlap region of the two plans. In this study, an automatic procedure was developed for field junction creation and lower extremities plan optimization.Methods
Ten patients treated with TMI-TMLI at our institution were selected retrospectively. The planning of the lower extremities was performed automatically. Target volume parameters (CTV_J‑V98% > 98%) at the junction region and several dose statistics (D98%, Dmean, and D2%) were compared between automatic and manual plans. The modulation complexity score (MCS) was used to assess plan complexity.Results
The automatic procedure required 60-90 min, depending on the case. All automatic plans achieved clinically acceptable dosimetric results (CTV_J‑V98% > 98%), with significant differences found at the junction region, where Dmean and D2% increased on average by 2.4% (p < 0.03) and 3.0% (p < 0.02), respectively. Similar plan complexity was observed (median MCS = 0.12). Since March 2022, the automatic procedure has been introduced in our clinic, reducing the TMI-TMLI simulation-to-delivery schedule by 2 days.Conclusion
The developed procedure allowed treatment planning of TMI-TMLI to be streamlined, increasing efficiency and standardization, preventing human errors, while maintaining the dosimetric plan quality and complexity of manual plans. Automated strategies can simplify the future adoption and clinical implementation of TMI-TMLI treatments in new centers.
SUBMITTER: Lambri N
PROVIDER: S-EPMC10033624 | biostudies-literature | 2023 Apr
REPOSITORIES: biostudies-literature
Lambri Nicola N Dei Damiano D Hernandez Victor V Castiglioni Isabella I Clerici Elena E Crespi Leonardo L De Philippis Chiara C Loiacono Daniele D Navarria Pierina P Reggiori Giacomo G Rusconi Roberto R Tomatis Stefano S Bramanti Stefania S Scorsetti Marta M Mancosu Pietro P
Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al] 20221103 4
<h4>Purpose</h4>Total marrow (and lymphoid) irradiation (TMI-TMLI) is limited by the couch travel range of modern linacs, which forces the treatment delivery to be split into two plans with opposite orientations: a head-first supine upper-body plan, and a feet-first supine lower extremities plan. A specific field junction is thus needed to obtain adequate target coverage in the overlap region of the two plans. In this study, an automatic procedure was developed for field junction creation and lo ...[more]