Unknown

Dataset Information

0

Non-continuous mobile electrocardiogram monitoring for post-transcatheter aortic valve replacement delayed conduction disorders put to the test.


ABSTRACT:

Aims

Permanent pacemaker implantation (PPM-I) remains nowadays the most important drawback of transcatheter aortic valve replacement (TAVR) procedure and the optimal strategy of delayed conduction disturbances (CDs) in these patients is unclear. The study aimed to validate an ambulatory electrocardiogram (ECG) monitoring through a 30 s spot ambulatory digital mobile ECG (AeECG), by using KardiaMobile-6L device in a 30-day period after TAVR procedure.

Methods and results

Between March 2021 and February 2022, we consecutively enrolled all patients undergoing TAVR procedure, except pacemaker (PM) carriers. At discharge, all patients were provided of a KardiaMobile-6L device and a spot digital ECG (eECG) recording 1 month schedule. Clinical and follow-up data were collected, and eECG schedule compliance and recording quality were explored. Among 151 patients without pre-existing PM, 23 were excluded for pre-discharge PPM-I, 18 failed the KardiaMobile-6L training phase, and 10 refused the device. Delayed CDs with a Class I/IIa indication for PPM-I occurred in eight patients (median 6 days). Delayed PPM-I vs. non-delayed PPM-I patients were more likely to have longer PR and QRS intervals at discharge. PR interval at discharge was the only independent predictor for delayed PPM-I at multivariate analysis. The overall eECG schedule compliance was 96.5%. None clinical adverse events CDs related were documented using this new AeECG monitoring modality.

Conclusion

A strategy of 30 s spot AeECG is safe and efficacious in delayed CDs monitoring after TAVR procedure with a very high eECG schedule level of compliance.

SUBMITTER: De Lucia R 

PROVIDER: S-EPMC10062351 | biostudies-literature | 2023 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Non-continuous mobile electrocardiogram monitoring for post-transcatheter aortic valve replacement delayed conduction disorders put to the test.

De Lucia Raffaele R   Giannini Cristina C   Parollo Matteo M   Barletta Valentina V   Costa Giulia G   Giannotti Santoro Mario M   Primerano Chiara C   Angelillis Marco M   De Carlo Marco M   Zucchelli Giulio G   Bongiorni Maria Grazia MG   Petronio Anna Sonia AS  

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology 20230301 3


<h4>Aims</h4>Permanent pacemaker implantation (PPM-I) remains nowadays the most important drawback of transcatheter aortic valve replacement (TAVR) procedure and the optimal strategy of delayed conduction disturbances (CDs) in these patients is unclear. The study aimed to validate an ambulatory electrocardiogram (ECG) monitoring through a 30 s spot ambulatory digital mobile ECG (AeECG), by using KardiaMobile-6L device in a 30-day period after TAVR procedure.<h4>Methods and results</h4>Between Ma  ...[more]

Similar Datasets

| S-EPMC8288806 | biostudies-literature
| S-EPMC8667767 | biostudies-literature
| S-EPMC4862065 | biostudies-other
| S-EPMC8642721 | biostudies-literature
| S-EPMC3709206 | biostudies-literature
| S-EPMC8183801 | biostudies-literature
| S-EPMC9912968 | biostudies-literature
| S-EPMC7237626 | biostudies-literature
| S-EPMC6790961 | biostudies-literature