4D-Flow Cardiovascular Magnetic Resonance Sequence for Aortic Assessment: Multi-Vendor and Multi-Magnetic Field Reproducibility in Healthy Volunteers.


ABSTRACT:

Objectives

Four-dimensional (4D) flow cardiac magnetic resonance (CMR) represents an emerging technique for non-invasive evaluation of the aortic flow. The aim of this study was to investigate a 4D-flow CMR sequence for the assessment of thoracic aorta comparing different vendors and different magnetic fields of MR scanner in fifteen healthy volunteers.

Methods

CMR was performed on three different MRI scanners: one at 1.5 T and two at 3 T. Flow parameters and planar wall shear stress (WSS) were extracted from six transversal planes along the full thoracic aorta by three operators. Inter-vendor comparability as well as scan-rescan, intra- and interobserver reproducibility were examined.

Results

A high heterogeneity was found in the comparisons for each operator and for each scanner in the six transversal planes analysis (Friedman rank-sum test; p-value ≤ 0.05). Among all, the most reproducible measures were extracted for the sinotubular junction plane and for the flow parameters.

Conclusions

Our results suggest that standardized procedures have to be defined to make more comparable and reproducible 4D-flow parameters and mainly, clinical impactfulness. Further studies on sequences development are needed to validate 4D-flow MRI assessment across vendors and magnetic fields also compared to a missing gold standard.

SUBMITTER: Punzo B 

PROVIDER: S-EPMC10141060 | biostudies-literature | 2023 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

4D-Flow Cardiovascular Magnetic Resonance Sequence for Aortic Assessment: Multi-Vendor and Multi-Magnetic Field Reproducibility in Healthy Volunteers.

Punzo Bruna B   Ranieri Brigida B   Tramontano Liberatore L   Affinito Ornella O   Franzese Monica M   Bossone Eduardo E   Saba Luca L   Cavaliere Carlo C   Cademartiri Filippo F  

Journal of clinical medicine 20230419 8


<h4>Objectives</h4>Four-dimensional (4D) flow cardiac magnetic resonance (CMR) represents an emerging technique for non-invasive evaluation of the aortic flow. The aim of this study was to investigate a 4D-flow CMR sequence for the assessment of thoracic aorta comparing different vendors and different magnetic fields of MR scanner in fifteen healthy volunteers.<h4>Methods</h4>CMR was performed on three different MRI scanners: one at 1.5 T and two at 3 T. Flow parameters and planar wall shear str  ...[more]

Similar Datasets

| S-EPMC5678712 | biostudies-literature
| S-EPMC7984392 | biostudies-literature
| S-EPMC11615600 | biostudies-literature
| S-EPMC7983287 | biostudies-literature
| S-EPMC6802342 | biostudies-literature
| S-EPMC11758371 | biostudies-literature
| S-EPMC8117268 | biostudies-literature
| S-EPMC9663893 | biostudies-literature
| S-EPMC5851552 | biostudies-literature
| S-EPMC5918697 | biostudies-literature