ABSTRACT: Objectives
Four-dimensional (4D) flow cardiac magnetic resonance (CMR) represents an emerging technique for non-invasive evaluation of the aortic flow. The aim of this study was to investigate a 4D-flow CMR sequence for the assessment of thoracic aorta comparing different vendors and different magnetic fields of MR scanner in fifteen healthy volunteers.Methods
CMR was performed on three different MRI scanners: one at 1.5 T and two at 3 T. Flow parameters and planar wall shear stress (WSS) were extracted from six transversal planes along the full thoracic aorta by three operators. Inter-vendor comparability as well as scan-rescan, intra- and interobserver reproducibility were examined.Results
A high heterogeneity was found in the comparisons for each operator and for each scanner in the six transversal planes analysis (Friedman rank-sum test; p-value ≤ 0.05). Among all, the most reproducible measures were extracted for the sinotubular junction plane and for the flow parameters.Conclusions
Our results suggest that standardized procedures have to be defined to make more comparable and reproducible 4D-flow parameters and mainly, clinical impactfulness. Further studies on sequences development are needed to validate 4D-flow MRI assessment across vendors and magnetic fields also compared to a missing gold standard.
SUBMITTER: Punzo B
PROVIDER: S-EPMC10141060 | biostudies-literature | 2023 Apr
REPOSITORIES: biostudies-literature
Punzo Bruna B Ranieri Brigida B Tramontano Liberatore L Affinito Ornella O Franzese Monica M Bossone Eduardo E Saba Luca L Cavaliere Carlo C Cademartiri Filippo F
Journal of clinical medicine 20230419 8
<h4>Objectives</h4>Four-dimensional (4D) flow cardiac magnetic resonance (CMR) represents an emerging technique for non-invasive evaluation of the aortic flow. The aim of this study was to investigate a 4D-flow CMR sequence for the assessment of thoracic aorta comparing different vendors and different magnetic fields of MR scanner in fifteen healthy volunteers.<h4>Methods</h4>CMR was performed on three different MRI scanners: one at 1.5 T and two at 3 T. Flow parameters and planar wall shear str ...[more]