Severe pulmonary valve insufficiency caused by transjugular cannulation of pulmonary artery for right ventricular assist device: diagnosis and surgical solution-a case report.


ABSTRACT: Implantation of a temporary percutaneous right ventricular assist device (RVAD) in patients with right heart failure after left ventricular assist device (LVAD) implantation is an established technique that may cause complications. We present a 60-year-old male patient who underwent urgent LVAD implantation. On the second postoperative day the patient developed acute right heart failure. We implanted a temporary percutaneous RVAD with two cannulas via the right internal jugular vein and the right femoral vein. Transesophageal echocardiography revealed severe pulmonary insufficiency. After performing re-sternotomy we anastomosed a prosthetic graft to the pulmonary trunk (PT), performed subxiphoid tunneling of the graft and replaced the transjugular outflow cannula. The pulmonary regurgitation caused by the percutaneous transvalvular cannula disappeared. In such case a direct anastomosis to the PT is the solution.

SUBMITTER: Wert L 

PROVIDER: S-EPMC10319755 | biostudies-literature | 2023 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Severe pulmonary valve insufficiency caused by transjugular cannulation of pulmonary artery for right ventricular assist device: diagnosis and surgical solution-a case report.

Wert Leonhard L   Lanmüller Pia P   Ott Sascha S   Mulzer Johanna J   Starck Christoph T CT   Falk Volkmar V   Potapov Evgenij V EV  

Journal of surgical case reports 20230703 7


Implantation of a temporary percutaneous right ventricular assist device (RVAD) in patients with right heart failure after left ventricular assist device (LVAD) implantation is an established technique that may cause complications. We present a 60-year-old male patient who underwent urgent LVAD implantation. On the second postoperative day the patient developed acute right heart failure. We implanted a temporary percutaneous RVAD with two cannulas via the right internal jugular vein and the righ  ...[more]

Similar Datasets

| S-EPMC7063611 | biostudies-literature
| S-EPMC10770093 | biostudies-literature
| S-EPMC9579997 | biostudies-literature
| S-EPMC5542743 | biostudies-other
| S-EPMC9619248 | biostudies-literature
| S-EPMC4295071 | biostudies-literature
| S-EPMC11345319 | biostudies-literature
| S-EPMC7089581 | biostudies-literature