Unknown

Dataset Information

0

Biomechanical and clinical evaluation of minimal invasive plate osteosynthesis for two-part clavicle shaft fractures.


ABSTRACT:

Background

Many surgical treatment methods exist for clavicle shaft fractures. A locking compression plate (LCP) fixation with three screws per fracture side is commonly used. For certain fractures a stabilization with 2 screws per side is potentially suitable, offering the advantage of reduced soft tissue approach, while avoiding the disadvantages of minimally-invasive nailing at the same time. This hypothesis was evaluated biomechanically and clinically.

Methods

Four treatment procedures were investigated biomechanically using composite human clavicle specimens. A load-to-failure test was performed using a three-point cantilever test. In group 1, a simple shaft fracture was simulated and stabilized with 2 screws per fracture side (5-hole LCP). In the second group 3 screws per side (7-hole LCP) were used. In group 3, a non-reduced fracture zone was simulated and treated with 3 screws per side (7-hole LCP). In group 4, an anatomically reduced fracture zone was simulated and treated with 3 screws per side (7-hole LCP). Furthermore 27 patients treated with a short plate and 2 screws per side (similar to group 1) were assessed after a minimum follow-up of 12 months (Constant and DASH Score).

Results

The maximum load-to-failure of group 1 was 367N. We observed the highest load-to-failure in group 2 with 497N and the lowest in group 3 with 90N. In group 4 a maximum load-to-failure of 298N could be evaluated. There was no significant difference in load-to-failure between the treatment of a simple clavicle fracture using 5- or 7-hole LCP (p = 0.121). However, we found a significant difference of load-to-failure between the simple and anatomically reduced fracture using a 7-hole plate (p = 0.014). The mean constant score of the surgically treated patients was 95 and the DASH score 3.0. Fracture consolidation was observed in 96.3%.

Conclusions

For certain non-fragmented and well interlocking 2-part fractures, a plate osteosynthesis fixed with only 2 screws per fracture side might offer sufficient biomechanical stability, better soft tissue preservation and comparable fusion rates compared to the operative treatment with 3 screws per side. However, the maximum load-to-failure of the 7-hole LCP was higher than of the 5-hole LCP, but this difference was not statistically significant.

Trial registration

Approval from the ethics committee of the Technical University of Dresden was retrospectively obtained (EK 588122019).

SUBMITTER: Schlußler A 

PROVIDER: S-EPMC10369786 | biostudies-literature | 2023 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Biomechanical and clinical evaluation of minimal invasive plate osteosynthesis for two-part clavicle shaft fractures.

Schlüßler Antonia A   Fehrenbacher Manuel M   Richter Richard Frank RF   Tille Eric E   Biewener Achim A   Nowotny Jörg J  

BMC musculoskeletal disorders 20230725 1


<h4>Background</h4>Many surgical treatment methods exist for clavicle shaft fractures. A locking compression plate (LCP) fixation with three screws per fracture side is commonly used. For certain fractures a stabilization with 2 screws per side is potentially suitable, offering the advantage of reduced soft tissue approach, while avoiding the disadvantages of minimally-invasive nailing at the same time. This hypothesis was evaluated biomechanically and clinically.<h4>Methods</h4>Four treatment p  ...[more]

Similar Datasets

| S-EPMC6831660 | biostudies-literature
| S-EPMC4447026 | biostudies-literature
| S-EPMC8079453 | biostudies-literature
| S-EPMC3750574 | biostudies-literature
| S-EPMC10693993 | biostudies-literature
| S-EPMC10351186 | biostudies-literature
| S-EPMC6205804 | biostudies-literature
| S-EPMC7677665 | biostudies-literature
| S-EPMC11837353 | biostudies-literature
| S-EPMC8732891 | biostudies-literature