Specific absorption rate (SAR) simulations for low-field (< 0.1 T) MRI systems.


ABSTRACT:

Objective

To simulate the magnetic and electric fields produced by RF coil geometries commonly used at low field. Based on these simulations, the specific absorption rate (SAR) efficiency can be derived to ensure safe operation even when using short RF pulses and high duty cycles.

Methods

Electromagnetic simulations were performed at four different field strengths between 0.05 and 0.1 T, corresponding to the lower and upper limits of current point-of-care (POC) neuroimaging systems. Transmit magnetic and electric fields, as well as transmit efficiency and SAR efficiency were simulated. The effects of a close-fitting shield on the EM fields were also assessed. SAR calculations were performed as a function of RF pulse length in turbo-spin echo (TSE) sequences.

Results

Simulations of RF coil characteristics and B1+ transmit efficiencies agreed well with corresponding experimentally determined parameters. Overall, the SAR efficiency was, as expected, higher at the lower frequencies studied, and many orders of magnitude greater than at conventional clinical field strengths. The tight-fitting transmit coil results in the highest SAR in the nose and skull, which are not thermally sensitive tissues. The calculated SAR efficiencies showed that only when 180° refocusing pulses of duration ~ 10 ms are used for TSE sequences does SAR need to be carefully considered.

Conclusion

This work presents a comprehensive overview of the transmit and SAR efficiencies for RF coils used for POC MRI neuroimaging. While SAR is not a problem for conventional sequences, the values derived here should be useful for RF intensive sequences such as T, and also demonstrate that if very short RF pulses are required then SAR calculations should be performed.

SUBMITTER: Parsa J 

PROVIDER: S-EPMC10386976 | biostudies-literature | 2023 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Specific absorption rate (SAR) simulations for low-field (< 0.1 T) MRI systems.

Parsa Javad J   Webb Andrew A  

Magma (New York, N.Y.) 20230318 3


<h4>Objective</h4>To simulate the magnetic and electric fields produced by RF coil geometries commonly used at low field. Based on these simulations, the specific absorption rate (SAR) efficiency can be derived to ensure safe operation even when using short RF pulses and high duty cycles.<h4>Methods</h4>Electromagnetic simulations were performed at four different field strengths between 0.05 and 0.1 T, corresponding to the lower and upper limits of current point-of-care (POC) neuroimaging system  ...[more]

Similar Datasets

| S-EPMC3871989 | biostudies-literature
| S-EPMC7396298 | biostudies-literature
| S-EPMC7722062 | biostudies-literature
| S-EPMC7815766 | biostudies-literature
| S-EPMC3946656 | biostudies-literature
| S-EPMC8166235 | biostudies-literature
| S-EPMC10952207 | biostudies-literature
| S-EPMC9328145 | biostudies-literature
| S-EPMC3897717 | biostudies-literature
| S-EPMC7807097 | biostudies-literature