[Formula: see text] Field inhomogeneity correction for qDESS [Formula: see text] mapping: application to rapid bilateral knee imaging.


ABSTRACT:

Purpose

[Formula: see text] mapping is a powerful tool for studying osteoarthritis (OA) changes and bilateral imaging may be useful in investigating the role of between-knee asymmetry in OA onset and progression. The quantitative double-echo in steady-state (qDESS) can provide fast simultaneous bilateral knee [Formula: see text] and high-resolution morphometry for cartilage and meniscus. The qDESS uses an analytical signal model to compute [Formula: see text] relaxometry maps, which require knowledge of the flip angle (FA). In the presence of [Formula: see text] inhomogeneities, inconsistencies between the nominal and actual FA can affect the accuracy of [Formula: see text] measurements. We propose a pixel-wise [Formula: see text] correction method for qDESS [Formula: see text] mapping exploiting an auxiliary [Formula: see text] map to compute the actual FA used in the model.

Methods

The technique was validated in a phantom and in vivo with simultaneous bilateral knee imaging. [Formula: see text] measurements of femoral cartilage (FC) of both knees of six healthy participants were repeated longitudinally to investigate the association between [Formula: see text] variation and [Formula: see text].

Results

The results showed that applying the [Formula: see text] correction mitigated [Formula: see text] variations that were driven by [Formula: see text] inhomogeneities. Specifically, [Formula: see text] left-right symmetry increased following the [Formula: see text] correction ([Formula: see text] = 0.74 > [Formula: see text] = 0.69). Without the [Formula: see text] correction, [Formula: see text] values showed a linear dependence with [Formula: see text]. The linear coefficient decreased using the [Formula: see text] correction (from 24.3 ± 1.6 ms to 4.1 ± 1.8) and the correlation was not statistically significant after the application of the Bonferroni correction (p value > 0.01).

Conclusion

The study showed that [Formula: see text] correction could mitigate variations driven by the sensitivity of the qDESS [Formula: see text] mapping method to [Formula: see text], therefore, increasing the sensitivity to detect real biological changes. The proposed method may improve the robustness of bilateral qDESS [Formula: see text] mapping, allowing for an accurate and more efficient evaluation of OA pathways and pathophysiology through longitudinal and cross-sectional studies.

SUBMITTER: Barbieri M 

PROVIDER: S-EPMC10524110 | biostudies-literature | 2023 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

B 1 Field inhomogeneity correction for qDESS T 2 mapping: application to rapid bilateral knee imaging.

Barbieri Marco M   Watkins Lauren E LE   Mazzoli Valentina V   Desai Arjun D AD   Rubin Elka E   Schmidt Andrew A   Gold Garry Evan GE   Hargreaves Brian Andrew BA   Chaudhari Akshay Sanjay AS   Kogan Feliks F  

Magma (New York, N.Y.) 20230504 5


<h4>Purpose</h4>T2 mapping is a powerful tool for studying osteoarthritis (OA) changes and bilateral imaging may be useful in investigating the role of between-knee asymmetry in OA onset and progression. The quantitative double-echo in steady-state (qDESS) can provide fast simultaneous bilateral knee T2 and high-resolution morphometry for cartilage and meniscus. The qDESS uses an analytical signal model to compute T2 relaxometry maps, which require knowledge of the flip angle (FA). In the presen  ...[more]

Similar Datasets

| S-EPMC4410410 | biostudies-other
| S-EPMC9701793 | biostudies-literature
| S-EPMC8494935 | biostudies-literature
| S-EPMC8595440 | biostudies-literature
| S-EPMC10539531 | biostudies-literature
| S-EPMC8671446 | biostudies-literature
| S-EPMC4424041 | biostudies-other
| S-EPMC7498615 | biostudies-literature
| S-EPMC11228484 | biostudies-literature
| S-EPMC8184963 | biostudies-literature