Quantitative 177Lu SPECT/CT imaging for personalized dosimetry using a ring-shaped CZT-based camera.


ABSTRACT:

Background

Dosimetry after radiopharmaceutical therapy with 177Lu (177Lu-RPT) relies on quantitative SPECT/CT imaging, for which suitable reconstruction protocols are required. In this study, we characterized for the first time the quantitative performance of a ring-shaped CZT-based camera using two different reconstruction algorithms: an ordered subset expectation maximization (OSEM) and a block sequential regularized expectation maximization (BSREM) combined with noise reduction regularization. This study lays the foundations for the definition of a reconstruction protocol enabling accurate dosimetry for patients treated with 177Lu-RPT.

Methods

A series of 177Lu-filled phantoms were acquired on a StarGuide™ (GE HealthCare), with energy and scatter windows centred at 208 (± 6%) keV and 185 (± 5%) keV, respectively. Images were reconstructed with the manufacturer implementations of OSEM (GE-OSEM) and BSREM (Q.Clear) algorithms, and various combinations of iterations and subsets. Additionally, the manufacturer-recommended Q.Clear-based reconstruction protocol was evaluated. Quantification accuracy, measured as the difference between the SPECT-based and the radionuclide calibrator-based activity, and noise were evaluated in a large cylinder. Recovery coefficients (RCs) and spatial resolution were assessed in a NEMA IEC phantom with sphere inserts. The reconstruction protocols considered suitable for clinical applications were tested on a cohort of patients treated with [177Lu]Lu-PSMA-I&T.

Results

The accuracy of the activity from the cylinder, although affected by septal penetration, was < 10% for all reconstructions. Both algorithms featured improved spatial resolution and higher RCs with increasing updates at the cost of noise build-up, but Q.Clear outperformed GE-OSEM in reducing noise accumulation. When the reconstruction parameters were carefully selected, similar values for noise (~0.15), spatial resolution (~1 cm) and RCs were found, irrespective of the reconstruction algorithm. Analogue results were found in patients.

Conclusions

Accurate activity quantification is possible when imaging 177Lu with StarGuide™. However, the impact of septal penetration requires further investigations. GE-OSEM is a valid alternative to the recommended Q.Clear reconstruction algorithm, featuring comparable performances assessed on phantoms and patients.

SUBMITTER: Danieli R 

PROVIDER: S-EPMC10584798 | biostudies-literature | 2023 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

Quantitative <sup>177</sup>Lu SPECT/CT imaging for personalized dosimetry using a ring-shaped CZT-based camera.

Danieli Rachele R   Stella Martina M   Leube Julian J   Tran-Gia Johannes J   Marin Clementine C   Uribe Carlos F CF   Vanderlinden Bruno B   Reynaert Nick N   Flamen Patrick P   Levillain Hugo H  

EJNMMI physics 20231018 1


<h4>Background</h4>Dosimetry after radiopharmaceutical therapy with <sup>177</sup>Lu (<sup>177</sup>Lu-RPT) relies on quantitative SPECT/CT imaging, for which suitable reconstruction protocols are required. In this study, we characterized for the first time the quantitative performance of a ring-shaped CZT-based camera using two different reconstruction algorithms: an ordered subset expectation maximization (OSEM) and a block sequential regularized expectation maximization (BSREM) combined with  ...[more]

Similar Datasets

| S-EPMC11043201 | biostudies-literature
| S-EPMC7343677 | biostudies-literature
| S-EPMC9081070 | biostudies-literature
| 2432699 | ecrin-mdr-crc
| S-EPMC7300169 | biostudies-literature
| S-EPMC9385894 | biostudies-literature
| S-EPMC10897098 | biostudies-literature
| S-EPMC10424930 | biostudies-literature
| S-EPMC10027798 | biostudies-literature
| S-EPMC9014836 | biostudies-literature