ABSTRACT: Objective
Assess the safety and efficacy of automated insulin delivery (AID) in adults with type 1 diabetes (T1D) at high risk for hypoglycemia.Research design and methods
Participants were 72 adults with T1D who used an insulin pump with Clarke Hypoglycemia Perception Awareness scale score >3 and/or had severe hypoglycemia during the previous 6 months confirmed by time below range (TBR; defined as sensor glucose [SG] reading <70 mg/dL) of at least 5% during 2 weeks of blinded continuous glucose monitoring (CGM). Parallel-arm, randomized trial (2:1) of AID (Tandem t:slim ×2 with Control-IQ technology) versus CGM and pump therapy for 12 weeks. The primary outcome was TBR change from baseline. Secondary outcomes included time in target range (TIR; 70-180 mg/dL), time above range (TAR), mean SG reading, and time with glucose level <54 mg/dL. An optional 12-week extension with AID was offered to all participants.Results
Compared with the sensor and pump (S&P), AID resulted in significant reduction of TBR by -3.7% (95% CI -4.8, -2.6), P < 0.001; an 8.6% increase in TIR (95% CI 5.2, 12.1), P < 0.001; and a -5.3% decrease in TAR (95% CI -87.7, -1.8), P = 0.004. Mean SG reading remained similar in the AID and S&P groups. During the 12-week extension, the effects of AID were sustained in the AID group and reproduced in the S&P group. Two severe hypoglycemic episodes occurred using AID.Conclusions
In adults with T1D at high risk for hypoglycemia, AID reduced the risk for hypoglycemia more than twofold, as quantified by TBR, while improving TIR and reducing hyperglycemia. Hence, AID is strongly recommended for this specific population.
SUBMITTER: Renard E
PROVIDER: S-EPMC10698222 | biostudies-literature | 2023 Dec
REPOSITORIES: biostudies-literature
Renard Eric E Joubert Michael M Villard Orianne O Dreves Bleuenn B Reznik Yves Y Farret Anne A Place Jerome J Breton Marc D MD Kovatchev Boris P BP
Diabetes care 20231201 12
<h4>Objective</h4>Assess the safety and efficacy of automated insulin delivery (AID) in adults with type 1 diabetes (T1D) at high risk for hypoglycemia.<h4>Research design and methods</h4>Participants were 72 adults with T1D who used an insulin pump with Clarke Hypoglycemia Perception Awareness scale score >3 and/or had severe hypoglycemia during the previous 6 months confirmed by time below range (TBR; defined as sensor glucose [SG] reading <70 mg/dL) of at least 5% during 2 weeks of blinded co ...[more]