Multicenter evaluation of the Selux Next-Generation Phenotyping antimicrobial susceptibility testing system.


ABSTRACT: The Selux Next-Generation Phenotyping (NGP) system (Charlestown, MA) is a new antimicrobial susceptibility testing system that utilizes two sequential assays performed on all wells of doubling dilution series to determine MICs. A multicenter evaluation of the performance of the Selux NGP system compared with reference broth microdilution was conducted following FDA recommendations and using FDA-defined breakpoints. A total of 2,488 clinical and challenge isolates were included; gram-negative isolates were tested against 24 antimicrobials, and gram-positive isolates were tested against 15 antimicrobials. Data is provided for all organism-antimicrobial combinations evaluated, including those that did and did not meet FDA performance requirements. Overall very major error and major error rates were less than 1% (31/3,805 and 107/15,606, respectively), essential agreement and categorical agreement were >95%, reproducibility was ≥95%, and the average time-to-result (from time of assay start to time of MIC result) was 5.65 hours.

SUBMITTER: Baker KR 

PROVIDER: S-EPMC10793272 | biostudies-literature | 2024 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Multicenter evaluation of the Selux Next-Generation Phenotyping antimicrobial susceptibility testing system.

Baker Kristin R KR   Flentie Kelly K   Spears Benjamin R BR   Mozharov Sergey S   Roberts Kristen K   El Ganbour Asmae A   Somers Mark M   Calkwood John J   Liu Jamie J   DaPonte Kayla K   Sam Nikitha N   Kaur Gurleen G   Chen Felicia F   Donato Jonathan J   Chao Alan A   Lewis Autumn A   Sherman Jingzi J   Mortimer Karen K   Harrington Amanda T AT   Traczewski Maria M   Carpenter Darcie D   Shortridge Dee D   Lindley Jill J   Diep Alexander A   Norton Emmet E   Green Matt M   Gajewski Joe J   Landrith Rebecca R   Nalubega Fatuma F   McCallum Justin J   Beiswenger Melissa M   Dolan Brittany B   Brennan Kathleen K   Carpenter Afton A   Vacic Aleksandar A   Flyer Alec N AN   Pierce Virginia M VM   Hooper David C DC   Lewis Ii James S JS   Stern Eric E  

Journal of clinical microbiology 20231205 1


The Selux Next-Generation Phenotyping (NGP) system (Charlestown, MA) is a new antimicrobial susceptibility testing system that utilizes two sequential assays performed on all wells of doubling dilution series to determine MICs. A multicenter evaluation of the performance of the Selux NGP system compared with reference broth microdilution was conducted following FDA recommendations and using FDA-defined breakpoints. A total of 2,488 clinical and challenge isolates were included; gram-negative iso  ...[more]

Similar Datasets

| S-EPMC5869823 | biostudies-other
| S-EPMC8106713 | biostudies-literature
| S-EPMC8212771 | biostudies-literature
| S-EPMC4938495 | biostudies-literature
| S-EPMC6535027 | biostudies-literature
| S-EPMC89554 | biostudies-literature
| S-EPMC8153291 | biostudies-literature
| S-EPMC8764967 | biostudies-literature
| S-EPMC5849110 | biostudies-literature
| S-EPMC3697729 | biostudies-literature