The relationships between lens diameter and ocular biometric parameters: an ultrasound biomicroscopy-based study.


ABSTRACT:

Purpose

This study aims to explore the relationships between lens diameter (LD) measured with ultrasound biomicroscopy (UBM) and ocular biometric parameters.

Methods

Ocular biometric parameters including axial length (AL), white-to-white distance (WTW), anterior chamber depth (ACD), lens thickness (LT) and anterior segment length (ASL) were measured with IOL-Master 700, and the direct measurement of LD was conducted through UBM (ArcScan Insight 100). Relationships between LD and ocular biometric parameters were then investigated. Eyes with AL ≥ 28 mm were defined as eyes with extreme myopia, and eyes with AL < 28 mm were defined as eyes without extreme myopia.

Results

A total of 194 eyes from 194 subjects were included. The mean LD was 9.58 ± 0.49 mm, ranging from 8.60 to 10.96 mm. According to univariate analysis, larger LD was associated with elder age, male gender, larger WTW, ACD and ASL (all p < 0.05). Meanwhile, the LD was positively correlated with AL in eyes without extreme myopia (p < 0.05), but not in eyes with extreme myopia (p > 0.05). Backward stepwise regressions revealed that a larger LD was associated with larger WTW, ASL and AL in eyes without extreme myopia (all p < 0.05), while ASL was the only significant variable in eyes with extreme myopia (p < 0.05).

Conclusion

Larger WTW, ASL and AL in eyes without extreme myopia, as well as longer ASL in eyes with extreme myopia indicated a larger LD, which provides guidance in personalized surgical choice and promises ideal visual outcomes.

SUBMITTER: Huang Z 

PROVIDER: S-EPMC10822951 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature

altmetric image

Publications

The relationships between lens diameter and ocular biometric parameters: an ultrasound biomicroscopy-based study.

Huang Zhiqian Z   Qi Jiao J   Cheng Kaiwen K   Liu Shuyu S   Zhang Keke K   Du Yu Y   Lu Yi Y   Zhu Xiangjia X  

Frontiers in medicine 20240115


<h4>Purpose</h4>This study aims to explore the relationships between lens diameter (LD) measured with ultrasound biomicroscopy (UBM) and ocular biometric parameters.<h4>Methods</h4>Ocular biometric parameters including axial length (AL), white-to-white distance (WTW), anterior chamber depth (ACD), lens thickness (LT) and anterior segment length (ASL) were measured with IOL-Master 700, and the direct measurement of LD was conducted through UBM (ArcScan Insight 100). Relationships between LD and o  ...[more]

Similar Datasets

| S-EPMC8882853 | biostudies-literature
| S-EPMC4816352 | biostudies-other
| S-EPMC10287596 | biostudies-literature
| S-EPMC10662558 | biostudies-literature
| S-EPMC10449110 | biostudies-literature
| S-EPMC8243629 | biostudies-literature
| S-EPMC9454160 | biostudies-literature
| S-EPMC6644274 | biostudies-literature
| S-EPMC10253972 | biostudies-literature
| S-EPMC4099472 | biostudies-literature