Human prostate MRI at ultrahigh-performance gradient: A feasibility study.


ABSTRACT:

Purpose

To demonstrate the technical feasibility and the value of ultrahigh-performance gradient in imaging the prostate in a 3T MRI system.

Methods

In this local institutional review board-approved study, prostate MRI was performed on 4 healthy men. Each subject was scanned in a prototype 3T MRI system with a 42-cm inner-diameter gradient coil that achieves a maximum gradient amplitude of 200 mT/m and slew rate of 500 T/m/s. PI-RADS V2.1-compliant axial T2 -weighted anatomical imaging and single-shot echo planar DWI at standard gradient of 70 mT/m and 150 T/m/s were obtained, followed by DWI at maximum performance (i.e., 200 mT/m and 500 T/m/s). In comparison to state-of-the-art clinical whole-body MRI systems, the high slew rate improved echo spacing from 1020 to 596 μs and, together with a high gradient amplitude for diffusion encoding, TE was reduced from 55 to 36 ms.

Results

In all 4 subjects (waist circumference = 81-91 cm, age = 45-65 years), no peripheral nerve stimulation sensation was reported during DWI. Reduced image distortion in the posterior peripheral zone prostate gland and higher signal intensity, such as in the surrounding muscle of high-gradient DWI, were noted.

Conclusion

Human prostate MRI at simultaneously high gradient amplitude of 200 mT/m and slew rate of 500 T/m/s is feasible, demonstrating that improved gradient performance can address image distortion and T2 decay-induced SNR issues for in vivo prostate imaging.

SUBMITTER: Zhu A 

PROVIDER: S-EPMC10841413 | biostudies-literature | 2024 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Purpose</h4>To demonstrate the technical feasibility and the value of ultrahigh-performance gradient in imaging the prostate in a 3T MRI system.<h4>Methods</h4>In this local institutional review board-approved study, prostate MRI was performed on 4 healthy men. Each subject was scanned in a prototype 3T MRI system with a 42-cm inner-diameter gradient coil that achieves a maximum gradient amplitude of 200 mT/m and slew rate of 500 T/m/s. PI-RADS V2.1-compliant axial T<sub>2</sub> -weighted an  ...[more]

Similar Datasets

| S-EPMC9755080 | biostudies-literature
| S-EPMC11607256 | biostudies-literature
| S-EPMC10104199 | biostudies-literature
| S-EPMC5556934 | biostudies-other
| S-EPMC8487192 | biostudies-literature
| S-EPMC7529434 | biostudies-literature
| S-EPMC6585650 | biostudies-literature
| S-EPMC11568117 | biostudies-literature
| S-EPMC8149725 | biostudies-literature
| S-EPMC3691850 | biostudies-other