Novel radiologic indices for stem type decision in total hip arthroplasty in patients with metaphyseo-diaphyseal mismatched Dorr A proximal femur.


ABSTRACT:

Background

In metaphyseo-diaphyseal (M-D) mismatched Dorr A femurs, it is difficult to achieve proper fixation with a type 1 stem. Proper interpretation of the geometry of the femur is integral at the preoperative stage in an M-D mismatched femur, but there has been a scarcity of studies on the radiologic indices. Therefore, we analyze the previous radiologic indices and suggest the novel ones for M-D mismatched femurs.

Methods

Our study was a retrospective review of preoperative radiographs of patients who underwent total hip arthroplasty with the smallest type 1 stem or with type 3 C stem at a single institution from July 2014 to March 2022. A Type 3 C stem was used when the smallest type 1 stem failed to achieve metaphyseal fixation. One hundred twenty-six patients were categorized into two main groups. Canal-flare index, canal-calcar ratio, modified morphological cortical index, and two novel indices (lesser trochanter-to-distal ratio-α and -β [LDR-α and -β]) were assessed on preoperative pelvic radiographs.

Results

Multivariate and ROC analysis demonstrated that high LDR-β (Exp[B]: 485.51, CI: 36.67-6427.97, p < 0.001) was associated with a more mismatched tendency group and had clinically acceptable discriminatory power (AUC: 0.765, CI: 0.675-0.855, p < 0.001) between the two cohorts.

Conclusion

Correct assessment of preoperative femoral morphology would be fundamental in the selection of a suitable stem. The ratio based on 3 cm below the lesser trochanter of the femur seemed crucial. We recommend evaluating the newly described radiological index preoperatively in M-D mismatched Dorr A femur for planning precisely and selecting a proper stem.

SUBMITTER: Lee HJ 

PROVIDER: S-EPMC10854119 | biostudies-literature | 2024 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Novel radiologic indices for stem type decision in total hip arthroplasty in patients with metaphyseo-diaphyseal mismatched Dorr A proximal femur.

Lee Han Jin HJ   Kim Hong Seok HS   Yoo Jeong Joon JJ  

BMC musculoskeletal disorders 20240209 1


<h4>Background</h4>In metaphyseo-diaphyseal (M-D) mismatched Dorr A femurs, it is difficult to achieve proper fixation with a type 1 stem. Proper interpretation of the geometry of the femur is integral at the preoperative stage in an M-D mismatched femur, but there has been a scarcity of studies on the radiologic indices. Therefore, we analyze the previous radiologic indices and suggest the novel ones for M-D mismatched femurs.<h4>Methods</h4>Our study was a retrospective review of preoperative  ...[more]

Similar Datasets

| S-EPMC11665467 | biostudies-literature
| S-EPMC9843827 | biostudies-literature
| S-EPMC9918064 | biostudies-literature
| S-EPMC6978188 | biostudies-literature
| S-EPMC8430435 | biostudies-literature
| S-EPMC5585817 | biostudies-literature
| S-EPMC11345508 | biostudies-literature
| S-EPMC8358097 | biostudies-literature
| S-EPMC10641077 | biostudies-literature
| S-EPMC8987314 | biostudies-literature