Safety of magnetic resonance imaging in patients with cardiac implantable electronic devices and abandoned or epicardial leads: a systematic review and meta-analysis.


ABSTRACT:

Aims

Persistent reluctance to perform magnetic resonance imaging (MRI) in patients with abandoned and/or epicardial leads of cardiac implantable electronic devices is related to in vitro studies reporting tip heating. While there is a plethora of data on the safety of MRI in conditional and non-conditional implantable devices, there is a clear lack of safety data in patients with abandoned and/or epicardial leads.

Methods and results

Relevant literature was identified in Medline and CINAHL using the key terms 'magnetic resonance imaging' AND 'abandoned leads' OR 'epicardial leads'. Secondary literature and cross-references were supplemented. For reporting guidance, the Preferred Reporting Items for Systematic reviews and Meta-Analyses 2020 was used. International Prospective Register of Systematic Reviews (PROSPERO) registration number 465530. Twenty-one publications with a total of 656 patients with 854 abandoned and/or epicardial leads and 929 MRI scans of different anatomical regions were included. No scan-related major adverse cardiac event was documented, although the possibility of under-reporting of critical events in the literature should be considered. Furthermore, no severe device dysfunction or severe arrhythmia was reported. Mainly transient lead parameter changes were observed in 2.8% in the subgroup of patients with functional epicardial leads. As a possible correlate of myocardial affection, subjective sensations occurred mainly in the subgroup with abandoned epicardial leads (4.0%), but no change in myocardial biomarkers was observed.

Conclusion

Existing publications did not report any relevant adverse events for MRI in patients with abandoned and/or epicardial leads if performed according to strict safety guidelines. However, a more rigorous risk-benefit calculation should be made for patients with epicardial leads.

SUBMITTER: Meier C 

PROVIDER: S-EPMC11200101 | biostudies-literature | 2024 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Safety of magnetic resonance imaging in patients with cardiac implantable electronic devices and abandoned or epicardial leads: a systematic review and meta-analysis.

Meier Claudia C   Israel Carsten C   Eisenblätter Michel M   Hoyer Annika A   Stoye Ferdinand Valentin FV   Yilmaz Ali A   Gielen Stephan S  

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology 20240601 6


<h4>Aims</h4>Persistent reluctance to perform magnetic resonance imaging (MRI) in patients with abandoned and/or epicardial leads of cardiac implantable electronic devices is related to in vitro studies reporting tip heating. While there is a plethora of data on the safety of MRI in conditional and non-conditional implantable devices, there is a clear lack of safety data in patients with abandoned and/or epicardial leads.<h4>Methods and results</h4>Relevant literature was identified in Medline a  ...[more]

Similar Datasets

| S-EPMC5557369 | biostudies-other
| S-EPMC9023390 | biostudies-literature
| S-EPMC9259370 | biostudies-literature
| S-EPMC9935018 | biostudies-literature
| S-EPMC5894885 | biostudies-literature
| S-EPMC11251694 | biostudies-literature
| S-EPMC10834167 | biostudies-literature
| S-EPMC6497358 | biostudies-literature
| S-EPMC5823604 | biostudies-literature
| S-EPMC10998729 | biostudies-literature