ABSTRACT: Background
Lower extremity peripheral artery disease frequently presents with calcifications which reduces the accuracy of computed tomography (CT) angiography, especially below-the-knee. Photon-counting detector (PCD)-CT offers improved spatial resolution and less calcium blooming. We aimed to identify the optimal reconstruction parameters for PCD-CT angiography of the lower legs.Methods
Tubes with different diameters (1-5 mm) were filled with different iodine concentrations and scanned in a water container. Images were reconstructed with 0.4 mm isotropic resolution using a quantitative kernel at all available sharpness levels (Qr36 to Qr76) and using different levels of quantum iterative reconstruction (QIR-2-4). Noise and image sharpness were determined for all reconstructions. Additionally, CT angiograms of 20 patients, reconstructed with a medium (Qr44), sharp (Qr60), and ultrasharp (Qr72) kernel at QIR-2-4, were evaluated by three readers assessing noise, delineation of plaques and vessel walls, and overall quality.Results
In the phantom study, increased kernel sharpness led to higher image noise (e.g., 16, 38, 77 HU for Qr44, Qr60, Qr72, and QIR-3). Image sharpness increased with increasing kernel sharpness, reaching a plateau at the medium-high level 60. Higher QIR levels decreased image noise (e.g., 51, 38, 25 HU at QIR-2-4 and Qr60) without reducing vessel sharpness. The qualitative in vivo results confirmed these findings: the sharp kernel (Qr60) with the highest QIR yielded the best overall quality.Conclusion
The combination of a sharpness level optimized reconstruction kernel (Qr60) and the highest QIR level yield the best image quality for PCD-CT angiography of the lower legs when reconstructed at 0.4-mm resolution.Relevance statement
Using high-resolution PCD-CT angiography with optimized reconstruction parameters might improve diagnostic accuracy and confidence in peripheral artery disease of the lower legs.Key points
Effective exploitation of the potential of PCD-CT angiography requires optimized reconstruction parameters. Too soft or too sharp reconstruction kernels reduce image quality. The highest level of quantum iterative reconstruction provides the best image quality.
SUBMITTER: Graafen D
PROVIDER: S-EPMC11294310 | biostudies-literature | 2024 Aug
REPOSITORIES: biostudies-literature
Graafen Dirk D Bart Willi W Halfmann Moritz C MC Müller Lukas L Hobohm Lukas L Yang Yang Y Neufang Achim A Espinola-Klein Christine C Pitton Michael B MB Kloeckner Roman R Varga-Szemes Akos A Emrich Tilman T
European radiology experimental 20240801 1
<h4>Background</h4>Lower extremity peripheral artery disease frequently presents with calcifications which reduces the accuracy of computed tomography (CT) angiography, especially below-the-knee. Photon-counting detector (PCD)-CT offers improved spatial resolution and less calcium blooming. We aimed to identify the optimal reconstruction parameters for PCD-CT angiography of the lower legs.<h4>Methods</h4>Tubes with different diameters (1-5 mm) were filled with different iodine concentrations and ...[more]