Atrioventricular Synchrony Delivered by a Dual-Chamber Leadless Pacemaker System.


ABSTRACT:

Background

A dual-chamber leadless pacemaker system has been designed for atrioventricular synchronous pacing using wireless, beat-to-beat, implant-to-implant (i2i) communication between distinct atrial and ventricular leadless pacemakers. The atrioventricular synchrony achieved across various ambulatory scenarios has yet to be systematically evaluated.

Methods

A prospective, single-arm, unblinded, multicenter, international clinical trial of the leadless pacemaker system was conducted in patients with a conventional dual-chamber pacing indication enrolled from February 2022 to March 2023. Leadless pacemaker systems were implanted, and 12-lead Holter electrocardiographic recordings were collected 3 months after implantation over various postures/activities: sitting, supine, left lateral recumbent, right lateral recumbent, standing, normal walk, and fast walk. An independent Holter core laboratory performed a manual adjudication of the percent of atrioventricular synchronous beats using the standard 300-millisecond PR interval limit. Atrium-to-ventricle and ventricle-to-atrium i2i communication success rates were also assessed. Post hoc summary statistics describing the relationships between atrioventricular synchrony and i2i success, posture/activity, implantation indication, atrioventricular event, and heart rate were calculated.

Results

In the evaluable population (n=384 of 464 enrolled [83%]; 61% male; age, 70 years; weight, 82 kg; 60% ejection fraction; 95% of beats evaluable), the mean atrioventricular synchrony of 98% of beats observed across all postures using the standard 300-millisecond limit was greater than both atrial-to-ventricular i2i (94%) and ventricular-to-atrial i2i (94%; P<0.001), exceeding both i2i values in 95% of patients. Atrioventricular synchrony was achieved in >95% of evaluable beats across all postures/activities, implantation indications, atrioventricular paced/sensed event combinations, and heart rate ranges (including >100 bpm).

Conclusions

This dual-chamber leadless pacemaker system demonstrated atrioventricular synchrony in 98% of evaluable beats at 3 months after implantation. Atrioventricular synchrony was maintained across postures/activities and remained robust for heart rates >100 bpm.

SUBMITTER: Ip JE 

PROVIDER: S-EPMC11305627 | biostudies-literature | 2024 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications


<h4>Background</h4>A dual-chamber leadless pacemaker system has been designed for atrioventricular synchronous pacing using wireless, beat-to-beat, implant-to-implant (i2i) communication between distinct atrial and ventricular leadless pacemakers. The atrioventricular synchrony achieved across various ambulatory scenarios has yet to be systematically evaluated.<h4>Methods</h4>A prospective, single-arm, unblinded, multicenter, international clinical trial of the leadless pacemaker system was cond  ...[more]

Similar Datasets

| S-EPMC10936979 | biostudies-literature
| S-EPMC8364773 | biostudies-literature
| S-EPMC11424992 | biostudies-literature
| S-EPMC7986103 | biostudies-literature
| S-EPMC11542626 | biostudies-literature
| S-EPMC8343438 | biostudies-literature
| S-EPMC10062361 | biostudies-literature
| S-EPMC8360010 | biostudies-literature