Transcatheter Extraction of a Migrated Left Atrial Appendage Occluder Device.


ABSTRACT: A 69-year-old man presented at 10 weeks postimplantation with a 31-mm Watchman FLX migrating into the left atrium. Due to incomplete left atrial appendage seal and embolization risk, transcatheter device extraction was performed without complications. Herein we describe the technique and procedural steps, using cardiac computed tomography and benchtop models to guide practice.

SUBMITTER: Wong G 

PROVIDER: S-EPMC11328763 | biostudies-literature | 2024 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Transcatheter Extraction of a Migrated Left Atrial Appendage Occluder Device.

Wong Gordon G   Aman Edris E   Kiaii Bob B   Rogers Jason H JH   Singh Gagan D GD  

JACC. Case reports 20240807 15


A 69-year-old man presented at 10 weeks postimplantation with a 31-mm Watchman FLX migrating into the left atrium. Due to incomplete left atrial appendage seal and embolization risk, transcatheter device extraction was performed without complications. Herein we describe the technique and procedural steps, using cardiac computed tomography and benchtop models to guide practice. ...[more]