Evaluation of agreement of IOP measurements by Tono-Vera tonometer to Goldmann applanation tonometry.


ABSTRACT:

Purpose

To evaluate the accuracy of the new Tono-Vera rebound tonometer (Reichert Inc, Buffalo, NY) compared to Goldmann Applanation Tonometry.

Methods

This prospective, observational, cross-sectional study was designed in accordance with ANSI Z80.10-2014 and ISO 8612-2009 guidelines for tonometer comparison. Intraocular Pressure (IOP) was measured by Goldmann Applanation and Tono-Vera on 160 eyes of 160 subjects. Corneal Astigmatism and Central Corneal Thickness were also measured. A single investigator (CN) conducted all measurements. The average of two measurements from each tonometer was used in the analysis. Bland-Altman plots, total least squares regression analysis, and simple linear regression were used to evaluate agreement between the tonometers.

Results

Average IOP values from Goldmann Applanation and Tono-Vera were not significantly different (19.17 and 19.03 respectively, p=0.40, paired t-test). The total least squares regression analysis indicated strong agreement between the two tonometers (slope +0.97, offset +0.49 mmHg, standard deviation 2.11 mmHg). There were 2 IOP measurement pairs that exceeded the ± 5 mmHg limits of agreement required in ANSI Z80.10-2014 and ISO 8612-2009, which is within the range of acceptability specified in the standards.

Conclusion

We evaluated IOP measurements by Tono-Vera Rebound Tonometer vs Goldmann Applanation Tonometry for eyes with a wide range of IOP values and found no statistically significant differences in the results. Tono-Vera meets the requirements of ANSI Z80.10-2014 and ISO 8612-2009, demonstrating accuracy comparable to Goldmann tonometry.

SUBMITTER: Niles CR 

PROVIDER: S-EPMC11362087 | biostudies-literature | 2024

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of agreement of IOP measurements by Tono-Vera tonometer to Goldmann applanation tonometry.

Niles Charles R CR   Crinzi Alexander R AR   Bonaventura Russell R   Taylor David A DA  

Frontiers in ophthalmology 20240816


<h4>Purpose</h4>To evaluate the accuracy of the new Tono-Vera rebound tonometer (Reichert Inc, Buffalo, NY) compared to Goldmann Applanation Tonometry.<h4>Methods</h4>This prospective, observational, cross-sectional study was designed in accordance with ANSI Z80.10-2014 and ISO 8612-2009 guidelines for tonometer comparison. Intraocular Pressure (IOP) was measured by Goldmann Applanation and Tono-Vera on 160 eyes of 160 subjects. Corneal Astigmatism and Central Corneal Thickness were also measure  ...[more]

Similar Datasets

| S-EPMC10412763 | biostudies-literature
| S-EPMC5292696 | biostudies-literature
| S-EPMC4290199 | biostudies-literature
| S-EPMC8247177 | biostudies-literature
| S-EPMC6857285 | biostudies-literature
| S-EPMC7606368 | biostudies-literature
| S-EPMC5540028 | biostudies-literature
| S-EPMC4212535 | biostudies-other
| S-EPMC3641509 | biostudies-literature
| S-EPMC7510959 | biostudies-literature