Evaluation of the Etest ESBL and the BD Phoenix, VITEK 1, and VITEK 2 automated instruments for detection of extended-spectrum beta-lactamases in multiresistant Escherichia coli and Klebsiella spp.


ABSTRACT: Seventy-four isolates of multiresistant Escherichia coli and Klebsiella spp. recovered during a 3-year period and 17 control strains with genotypically identified beta-lactamases were tested for the production of extended-spectrum beta-lactamases (ESBLs) by using the Etest and the VITEK 1, VITEK 2, and Phoenix automated instruments. The use of the Etest was evaluated by investigating its accuracy in detecting the ESBLs of the control strains and by comparing interpretation results of laboratory technicians and experts. The accuracy of the Etest was 94%. With the Etest as the reference for the clinical strains and the genotype as the reference for the control strains, the automated instruments detected the ESBLs with accuracies of 78% (VITEK 2), 83% (VITEK 1), and 89% (Phoenix). No signific