In vivo micro-CT scanning of a rabbit distal femur: repeatability and reproducibility.
Ontology highlight
ABSTRACT: Before in vivo micro-CT scanning can be used to investigate femoral trabecular microarchitecture over time in rabbits, its repeatability and reproducibility must be demonstrated. To accomplish this, both distal femurs of two 6-month-old New Zealand white rabbits were scanned five times each in 1 day under different conditions (repeatability). Scanning was done at 28 microm isotropic voxel size to produce five image stacks of each femur. Three operators then followed a standard image processing protocol (reproducibility) to isolate two separate cubes from each anterior femoral condyle [total n = (8 cube sites)(5 scans)(3 operators) = 120]. Bone volume fraction (BV/TV) of the eight different cube sites (sample) ranged from 0.408 to 0.501 (mean: 0.453); trabecular thickness (Tb.Th) ranged fro
SUBMITTER: Voor MJ
PROVIDER: S-EPMC2244798 | biostudies-literature | 2008
REPOSITORIES: biostudies-literature
ACCESS DATA