Single-point mutations causing more than 100-fold underestimation of human immunodeficiency virus type 1 (HIV-1) load with the Cobas TaqMan HIV-1 real-time PCR assay.


ABSTRACT: Systematic sequence analysis of human immunodeficiency virus type 1 (HIV-1) variants with RNA levels underestimated by the Cobas TaqMan HIV-1 assay demonstrated that mutations at a single position of the downstream primer can lead to the underestimation of HIV-1 RNA levels by more than 2 log and to false-negative results in minipool screening of blood donors. Mutations at this position are found in about 2% of all HIV-1 M gag sequences.

SUBMITTER: Korn K 

PROVIDER: S-EPMC2668364 | biostudies-literature | 2009 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Single-point mutations causing more than 100-fold underestimation of human immunodeficiency virus type 1 (HIV-1) load with the Cobas TaqMan HIV-1 real-time PCR assay.

Korn Klaus K   Weissbrich Benedikt B   Henke-Gendo Cornelia C   Heim Albert A   Jauer Christin Mariel CM   Taylor Ninon N   Eberle Josef J  

Journal of clinical microbiology 20090204 4


Systematic sequence analysis of human immunodeficiency virus type 1 (HIV-1) variants with RNA levels underestimated by the Cobas TaqMan HIV-1 assay demonstrated that mutations at a single position of the downstream primer can lead to the underestimation of HIV-1 RNA levels by more than 2 log and to false-negative results in minipool screening of blood donors. Mutations at this position are found in about 2% of all HIV-1 M gag sequences. ...[more]