ABSTRACT: Objective
To compare the sensitivity and specificity of luciferase immunoprecipitation (LIPS) with radioimmunoprecipitation (RIP) for the measurement of autoantibodies to the type 1 diabetes autoantigens glutamic acid decarboxylase 65 (GAD65) and insulinoma-associated protein (IA)-2beta.Research design and methods
Sera from 49 type 1 diabetic patients and 100 nondiabetic control subjects from Diabetes Antibody Standardization Program 2007 were used to screen for autoantibodies to GAD65. An additional 200 type 1 diabetic patients and 200 nondiabetic control subjects were used to validate the GAD65 results and screen for autoantibodies to IA-2beta.Results
LIPS showed equal sensitivity and specificity to RIP for detecting autoantibodies to GAD65 and IA-2beta. Receiver-operating characteristic analysis revealed that the detection of autoantibodies to GAD65 and IA-2beta by LIPS and RIP were not statistically different.Conclusions
The LIPS assay does not require the use of radioisotopes or in vitro transcription/translation and is a practical alternative at the clinical level for the RIP assay.
SUBMITTER: Burbelo PD
PROVIDER: S-EPMC2845022 | biostudies-literature | 2010 Apr
REPOSITORIES: biostudies-literature
Burbelo Peter D PD Hirai Hiroki H Issa Alexandra T AT Kingman Albert A Lernmark Ake A Ivarsson Sten-A SA Notkins Abner L AL Iadarola Michael J MJ
Diabetes care 20100119 4
<h4>Objective</h4>To compare the sensitivity and specificity of luciferase immunoprecipitation (LIPS) with radioimmunoprecipitation (RIP) for the measurement of autoantibodies to the type 1 diabetes autoantigens glutamic acid decarboxylase 65 (GAD65) and insulinoma-associated protein (IA)-2beta.<h4>Research design and methods</h4>Sera from 49 type 1 diabetic patients and 100 nondiabetic control subjects from Diabetes Antibody Standardization Program 2007 were used to screen for autoantibodies to ...[more]