An inter-laboratory comparison of urinary 3-hydroxypropylmercapturic acid measurement demonstrates good reproducibility between laboratories.


ABSTRACT:

Background

Biomarkers have been used extensively in clinical studies to assess toxicant exposure in smokers and non-smokers and have recently been used in the evaluation of novel tobacco products. The urinary metabolite 3-HPMA, a metabolite of the major tobacco smoke toxicity contributor acrolein, is one example of a biomarker used to measure exposure to tobacco smoke. A number of laboratories have developed liquid chromatography with tandem mass spectrometry (LC-MS/MS) based methods to measure urinary 3-HPMA; however, it is unclear to what extent the data obtained by these different laboratories are comparable.

Findings

This report describes an inter-laboratory comparison carried out to evaluate the comparability of 3-HPMA measurement between four laboratories. A common set of spiked and authentic smoker and non-smoker urine samples were used. Each laboratory used their in-house LC-MS/MS method and a common internal standard. A comparison of the repeatability ('r'), reproducibility ('R'), and coefficient of variation for 3-HPMA demonstrated that within-laboratory variation was consistently lower than between-laboratory variation. The average inter-laboratory coefficient of variation was 7% for fortified urine samples and 16.2% for authentic urine samples. Together, this represents an inter-laboratory variation of 12.2%.

Conclusion

The results from this first inter-laboratory comparison for the measurement of 3-HPMA in urine demonstrate a reasonably good consensus between laboratories. However, some consistent measurement biases were still observed between laboratories, suggesting that additional work may be required to further reduce the inter-laboratory coefficient of variation.

SUBMITTER: Minet E 

PROVIDER: S-EPMC3208382 | biostudies-literature | 2011 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

An inter-laboratory comparison of urinary 3-hydroxypropylmercapturic acid measurement demonstrates good reproducibility between laboratories.

Minet Emmanuel E   Errington Graham G   Scherer Gerhard G   Newland Kirk K   Sharifi Mehran M   Bailey Brian B   McEwan Mike M   Cheung Francis F  

BMC research notes 20111010


<h4>Background</h4>Biomarkers have been used extensively in clinical studies to assess toxicant exposure in smokers and non-smokers and have recently been used in the evaluation of novel tobacco products. The urinary metabolite 3-HPMA, a metabolite of the major tobacco smoke toxicity contributor acrolein, is one example of a biomarker used to measure exposure to tobacco smoke. A number of laboratories have developed liquid chromatography with tandem mass spectrometry (LC-MS/MS) based methods to  ...[more]

Similar Datasets

| S-EPMC7229280 | biostudies-literature
| S-EPMC10508307 | biostudies-literature
| S-EPMC7552071 | biostudies-literature
| S-EPMC6608591 | biostudies-literature
| S-EPMC5310769 | biostudies-literature
| S-EPMC9038857 | biostudies-literature
| S-EPMC6918145 | biostudies-literature
| S-EPMC1914270 | biostudies-other
| S-EPMC7010240 | biostudies-literature
| S-EPMC3085423 | biostudies-literature