Unknown

Dataset Information

0

Introduction of new industry standards for cardiac implantable electronic devices: balancing benefits and unexpected risks.


ABSTRACT: The DF-4 connector is a novel industry standard for the connection of a defibrillator lead to the generator. It aims at reducing the bulk created by two or three pins at the proximal end of the defibrillator lead and its corresponding ports at the header of the device. Having only one connection port between the lead and the device reduces the material in the pocket, the risk of lead-to-port mismatch, may lower the risk of lead abrasion, and probably makes the implantation procedure a little easier since only one set screw is required. However, all these conceived benefits are related to convenience rather than to a medical need. After the recent experiences with the possible negative clinical impact of 'minor' changes like simply downsizing a defibrillator lead, a word of caution is warranted. The lead is the weakest part of the defibrillator system, complex in design and undergoing constant stress through movement. It is very hard to predict which issues may evolve over time with the changes in lead design. Does the perceived benefit really outweigh an unpredictable risk in a sensitive medical product like a defibrillator? This article tries to address the possible issues of the new spring contacts instead of set screws, the proximity of the low- and high-voltage connections as well as the inability of adding a pace/sense or an additional shock lead without a special adaptor, and advocates a measured speed in the introduction of this technology.

SUBMITTER: Sticherling C 

PROVIDER: S-EPMC3404557 | biostudies-literature | 2012 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Introduction of new industry standards for cardiac implantable electronic devices: balancing benefits and unexpected risks.

Sticherling Christian C   Burri Haran H  

Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology 20120301 8


The DF-4 connector is a novel industry standard for the connection of a defibrillator lead to the generator. It aims at reducing the bulk created by two or three pins at the proximal end of the defibrillator lead and its corresponding ports at the header of the device. Having only one connection port between the lead and the device reduces the material in the pocket, the risk of lead-to-port mismatch, may lower the risk of lead abrasion, and probably makes the implantation procedure a little eas  ...[more]

Similar Datasets

| S-EPMC5517255 | biostudies-other
| S-EPMC5837548 | biostudies-literature
| S-EPMC4974896 | biostudies-other
| S-EPMC9247999 | biostudies-literature
| S-EPMC5127493 | biostudies-other
| S-EPMC3616683 | biostudies-other
| S-EPMC2748105 | biostudies-literature
| S-EPMC4711570 | biostudies-literature
| S-EPMC3417886 | biostudies-other
| S-EPMC6872914 | biostudies-literature