Volumetric real-time imaging using a CMUT ring array.


ABSTRACT: A ring array provides a very suitable geometry for forward-looking volumetric intracardiac and intravascular ultrasound imaging. We fabricated an annular 64-element capacitive micromachined ultrasonic transducer (CMUT) array featuring a 10-MHz operating frequency and a 1.27-mm outer radius. A custom software suite was developed to run on a PC-based imaging system for real-time imaging using this device. This paper presents simulated and experimental imaging results for the described CMUT ring array. Three different imaging methods--flash, classic phased array (CPA), and synthetic phased array (SPA)--were used in the study. For SPA imaging, two techniques to improve the image quality--Hadamard coding and aperture weighting--were also applied. The results show that SPA with Hadamard coding and aperture weighting is a good option for ring-array imaging. Compared with CPA, it achieves better image resolution and comparable signal-to-noise ratio at a much faster image acquisition rate. Using this method, a fast frame rate of up to 463 volumes per second is achievable if limited only by the ultrasound time of flight; with the described system we reconstructed three cross-sectional images in real-time at 10 frames per second, which was limited by the computation time in synthetic beamforming.

SUBMITTER: Choe JW 

PROVIDER: S-EPMC3415797 | biostudies-literature | 2012 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Volumetric real-time imaging using a CMUT ring array.

Choe Jung Woo JW   Oralkan Ömer Ö   Nikoozadeh Amin A   Gencel Mustafa M   Stephens Douglas N DN   O'Donnell Matthew M   Sahn David J DJ   Khuri-Yakub Butrus T BT  

IEEE transactions on ultrasonics, ferroelectrics, and frequency control 20120601 6


A ring array provides a very suitable geometry for forward-looking volumetric intracardiac and intravascular ultrasound imaging. We fabricated an annular 64-element capacitive micromachined ultrasonic transducer (CMUT) array featuring a 10-MHz operating frequency and a 1.27-mm outer radius. A custom software suite was developed to run on a PC-based imaging system for real-time imaging using this device. This paper presents simulated and experimental imaging results for the described CMUT ring ar  ...[more]

Similar Datasets

| S-EPMC4019434 | biostudies-literature
| S-EPMC4074521 | biostudies-literature
| S-EPMC8585549 | biostudies-literature
| S-EPMC6444985 | biostudies-literature
| S-EPMC9388629 | biostudies-literature
| S-EPMC10521654 | biostudies-literature
| S-EPMC3089374 | biostudies-literature
| S-EPMC8758712 | biostudies-literature
| S-EPMC7118155 | biostudies-literature
| S-EPMC10314963 | biostudies-literature