Accuracy of a dual path platform (DPP) assay for the rapid point-of-care diagnosis of human leptospirosis.


ABSTRACT:

Background

Diagnosis of leptospirosis by the gold standard serologic assay, the microscopic agglutination test (MAT), requires paired sera and is not widely available. We developed a rapid assay using immunodominant Leptospira immunoglobulin-like (Lig) proteins in a Dual Path Platform (DPP). This study aimed to evaluate the assay's diagnostic performance in the setting of urban transmission.

Methodology

We determined test sensitivity using 446 acute and convalescent sera from MAT-confirmed case-patients with severe or mild leptospirosis in Brazil. We assessed test specificity using 677 sera from the following groups: healthy residents of a Brazilian slum with endemic transmission, febrile outpatients from the same slum, healthy blood donors, and patients with dengue, hepatit