Unknown

Dataset Information

0

LC-MS-MS Measurements of Urinary Creatinine and the Application of Creatinine Normalization Technique on Cotinine in Smokers' 24 Hour Urine.


ABSTRACT: A simple and sensitive high performance liquid chromatography-tandem mass spectrometry (HPLC-ESI-MS-MS) method was developed and validated for the quantification of creatinine in human urine. The analysis was carried out on an Agilent Zorbax Eclipse XDB-C18 column (2.1 × 150?mm, 3.5??m). The mobile phase was 0.1% formic acid in water and 0.1% formic acid in acetonitrile (50/50, v/v). Linear calibration curves were obtained in the concentration range of 1-2000.0?ng/mL, with a lower limit of quantification of 0.99?ng/mL. The intra- and interday precision (RSD) values were below 3%. The method was successfully applied to a bioequivalence study of creatinine in Chinese smokers and nonsmokers. The total cotinine in 24?h urine and cotinine?:?creatinine ratio were also positively associated (Pearson R = 0.942, P < 0.0001). However, cotinine?:?creatinine ratio varied significantly across smoking groups for the difference of individual. 24?h urinary cotinine was more appropriate for expressing correlation with tar than cotinine?:?creatinine ratio.

SUBMITTER: Hou H 

PROVIDER: S-EPMC3503330 | biostudies-literature | 2012

REPOSITORIES: biostudies-literature

altmetric image

Publications

LC-MS-MS Measurements of Urinary Creatinine and the Application of Creatinine Normalization Technique on Cotinine in Smokers' 24 Hour Urine.

Hou Hongwei H   Xiong Wei W   Zhang Xiaotao X   Song Dongkui D   Tang Gangling G   Hu Qingyuan Q  

Journal of analytical methods in chemistry 20121111


A simple and sensitive high performance liquid chromatography-tandem mass spectrometry (HPLC-ESI-MS-MS) method was developed and validated for the quantification of creatinine in human urine. The analysis was carried out on an Agilent Zorbax Eclipse XDB-C18 column (2.1 × 150 mm, 3.5 μm). The mobile phase was 0.1% formic acid in water and 0.1% formic acid in acetonitrile (50/50, v/v). Linear calibration curves were obtained in the concentration range of 1-2000.0 ng/mL, with a lower limit of quant  ...[more]

Similar Datasets

| S-EPMC7146571 | biostudies-literature
| S-EPMC6401114 | biostudies-literature
| S-EPMC6738107 | biostudies-literature
| S-EPMC3700702 | biostudies-literature
| S-EPMC4514740 | biostudies-literature
| S-EPMC5911189 | biostudies-literature
| S-EPMC4073156 | biostudies-literature
| S-EPMC8575727 | biostudies-literature
| S-EPMC5733879 | biostudies-literature
| S-EPMC4094486 | biostudies-literature