Are fluorescence-detected sedimentation velocity data reliable?


ABSTRACT: Sedimentation velocity analytical ultracentrifugation is a classical biophysical technique that is commonly used to analyze the size, shape, and interactions of biological macromolecules in solution. Fluorescence detection provides enhanced sensitivity and selectivity relative to the standard absorption and refractrometric detectors, but data acquisition is more complex and can be subject to interference from several photophysical effects. Here, we describe methods to configure sedimentation velocity measurements using fluorescence detection and evaluate the performance of the fluorescence optical system. The fluorescence detector output is linear over a concentration range of at least 1 to 500nM fluorescein and Alexa Fluor 488. At high concentrations, deviations from linearity can be attributed to the inner filter effect. A duplex DNA labeled with Alexa Fluor 488 was used as a standard to compare sedimentation coefficients obtained using fluorescence and absorbance detectors. Within error, the sedimentation coefficients agree. Thus, the fluorescence detector is capable of providing precise and accurate sedimentation velocity results that are consistent with measurements performed using conventional absorption optics, provided the data are collected at appropriate sample concentrations and the optics are configured correctly.

SUBMITTER: Lyons DF 

PROVIDER: S-EPMC3640771 | biostudies-literature | 2013 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Are fluorescence-detected sedimentation velocity data reliable?

Lyons Daniel F DF   Lary Jeffrey W JW   Husain Bushra B   Correia John J JJ   Cole James L JL  

Analytical biochemistry 20130307 2


Sedimentation velocity analytical ultracentrifugation is a classical biophysical technique that is commonly used to analyze the size, shape, and interactions of biological macromolecules in solution. Fluorescence detection provides enhanced sensitivity and selectivity relative to the standard absorption and refractrometric detectors, but data acquisition is more complex and can be subject to interference from several photophysical effects. Here, we describe methods to configure sedimentation vel  ...[more]

Similar Datasets

| S-EPMC7466938 | biostudies-literature
| S-EPMC3988680 | biostudies-literature
| S-EPMC4165462 | biostudies-literature
| S-EPMC6768728 | biostudies-literature
| S-EPMC7701085 | biostudies-literature
| S-EPMC3799624 | biostudies-literature
| S-EPMC5639329 | biostudies-literature
| S-EPMC5765866 | biostudies-literature
| S-EPMC3058617 | biostudies-literature
| S-EPMC6158090 | biostudies-literature