Recorded scan times can limit the accuracy of sedimentation coefficients in analytical ultracentrifugation.


ABSTRACT: We report systematic and large inaccuracies in the recorded elapsed time in data files from the analytical ultracentrifuge, leading to overestimates of the sedimentation coefficients of up to 10%. This far exceeds previously considered factors contributing to the uncertainty in this parameter and has significant ramifications for derived parameters such as hydrodynamic shape and molar mass estimates. The source of this error is currently unknown, but we found it to be quantitatively consistent across different instruments, increasing with rotor speed. Furthermore, its occurrence appears to correlate with the use of the latest data acquisition software from the manufacturer, in use in some of our laboratories for nearly 2 years. Many of the recently published sedimentation coefficients may need to be reexamined. The problem can be easily recognized by comparing the file timestamps provided by the operating system with the elapsed scan times recorded within the data files. Therefore, we implemented a routine in SEDFIT that can automatically examine the data files, alert the user to significant discrepancies, and correct the scan times accordingly. This eliminates errors in the recorded scan times.

SUBMITTER: Zhao H 

PROVIDER: S-EPMC3676908 | biostudies-literature | 2013 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Recorded scan times can limit the accuracy of sedimentation coefficients in analytical ultracentrifugation.

Zhao Huaying H   Ghirlando Rodolfo R   Piszczek Grzegorz G   Curth Ute U   Brautigam Chad A CA   Schuck Peter P  

Analytical biochemistry 20130228 1


We report systematic and large inaccuracies in the recorded elapsed time in data files from the analytical ultracentrifuge, leading to overestimates of the sedimentation coefficients of up to 10%. This far exceeds previously considered factors contributing to the uncertainty in this parameter and has significant ramifications for derived parameters such as hydrodynamic shape and molar mass estimates. The source of this error is currently unknown, but we found it to be quantitatively consistent a  ...[more]

Similar Datasets

| S-EPMC5899733 | biostudies-literature
| S-EPMC4547334 | biostudies-literature
| S-EPMC4805870 | biostudies-other
| S-EPMC3799624 | biostudies-literature
| S-EPMC5809250 | biostudies-literature
| S-EPMC3866193 | biostudies-literature
| S-EPMC7398841 | biostudies-literature
| S-EPMC5899701 | biostudies-literature
| S-EPMC4440767 | biostudies-literature
| S-EPMC6066226 | biostudies-literature