Nicked-sleeve interface for two-dimensional capillary electrophoresis.


ABSTRACT: We report an improved interface for two-dimensional capillary electrophoresis. This interface is based on capillary tubing and a Plexiglas chip, both of which were milled using a micro-dicing saw. The interface was evaluated and compared to a traditional interface design for both pseudo one-dimensional and two-dimensional capillary electrophoresis. We observe less than 70% transfer efficiency for the traditional design and greater than 90% transfer efficiency with this new interface.

SUBMITTER: Flaherty RJ 

PROVIDER: S-EPMC3697014 | biostudies-literature | 2013 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Nicked-sleeve interface for two-dimensional capillary electrophoresis.

Flaherty Ryan J RJ   Huge Bonnie J BJ   Bruce Spencer M SM   Dada Oluwatosin O OO   Dovichi Norman J NJ  

The Analyst 20130701 13


We report an improved interface for two-dimensional capillary electrophoresis. This interface is based on capillary tubing and a Plexiglas chip, both of which were milled using a micro-dicing saw. The interface was evaluated and compared to a traditional interface design for both pseudo one-dimensional and two-dimensional capillary electrophoresis. We observe less than 70% transfer efficiency for the traditional design and greater than 90% transfer efficiency with this new interface. ...[more]

Similar Datasets

| S-EPMC3437655 | biostudies-literature
| S-EPMC3059363 | biostudies-literature
| S-EPMC3038187 | biostudies-literature
| S-EPMC10416412 | biostudies-literature
| S-EPMC5217355 | biostudies-literature
| S-EPMC4757711 | biostudies-literature
| S-EPMC8761713 | biostudies-literature
| S-EPMC3865863 | biostudies-literature
| S-EPMC3714212 | biostudies-literature
| S-EPMC9562008 | biostudies-literature