Unknown

Dataset Information

0

Myocardial T1 and T2 mapping at 3 T: reference values, influencing factors and implications.


ABSTRACT:

Background

Myocardial T1 and T2 mapping using cardiovascular magnetic resonance (CMR) are promising to improve tissue characterization and early disease detection. This study aimed at analyzing the feasibility of T1 and T2 mapping at 3 T and providing reference values.

Methods

Sixty healthy volunteers (30 males/females, each 20 from 20-39 years, 40-59 years, 60-80 years) underwent left-ventricular T1 and T2 mapping in 3 short-axis slices at 3 T. For T2 mapping, 3 single-shot steady-state free precession (SSFP) images with different T2 preparation times were acquired. For T1 mapping, modified Look-Locker inversion recovery technique with 11 single shot SSFP images was used before and after injection of gadolinium contrast. T1 and T2 relaxation times were quantified for each slice and each myocardial segment.

Results

Mean T2 and T1 (pre-/post-contrast) times were: 44.1 ms/1157.1 ms/427.3 ms (base), 45.1 ms/1158.7 ms/411.2 ms (middle), 46.9 ms/1180.6 ms/399.7 ms (apex). T2 and pre-contrast T1 increased from base to apex, post-contrast T1 decreased. Relevant inter-subject variability was apparent (scatter factor 1.08/1.05/1.11 for T2/pre-contrast T1/post-contrast T1). T2 and post-contrast T1 were influenced by heart rate (p?ConclusionsMyocardial T2 and T1 reference values for the specific CMR setting are provided. The diagnostic impact of the high inter-subject variability of T2 and T1 relaxation times requires further investigation.

SUBMITTER: von Knobelsdorff-Brenkenhoff F 

PROVIDER: S-EPMC3702448 | biostudies-literature | 2013 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Myocardial T1 and T2 mapping at 3 T: reference values, influencing factors and implications.

von Knobelsdorff-Brenkenhoff Florian F   Prothmann Marcel M   Dieringer Matthias A MA   Wassmuth Ralf R   Greiser Andreas A   Schwenke Carsten C   Niendorf Thoralf T   Schulz-Menger Jeanette J  

Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance 20130618


<h4>Background</h4>Myocardial T1 and T2 mapping using cardiovascular magnetic resonance (CMR) are promising to improve tissue characterization and early disease detection. This study aimed at analyzing the feasibility of T1 and T2 mapping at 3 T and providing reference values.<h4>Methods</h4>Sixty healthy volunteers (30 males/females, each 20 from 20-39 years, 40-59 years, 60-80 years) underwent left-ventricular T1 and T2 mapping in 3 short-axis slices at 3 T. For T2 mapping, 3 single-shot stead  ...[more]

Similar Datasets

| S-EPMC4811754 | biostudies-literature
| S-EPMC10652608 | biostudies-literature
| S-EPMC6817390 | biostudies-literature
| S-EPMC5740192 | biostudies-literature
| S-EPMC8442989 | biostudies-literature
| S-EPMC9306903 | biostudies-literature
| S-EPMC5563030 | biostudies-other
| S-EPMC4376813 | biostudies-literature
| S-EPMC3424120 | biostudies-literature
| S-EPMC9205876 | biostudies-literature