Evaluation of slice accelerations using multiband echo planar imaging at 3 T.


ABSTRACT: We evaluate residual aliasing among simultaneously excited and acquired slices in slice accelerated multiband (MB) echo planar imaging (EPI). No in-plane accelerations were used in order to maximize and evaluate achievable slice acceleration factors at 3 T. We propose a novel leakage (L-) factor to quantify the effects of signal leakage between simultaneously acquired slices. With a standard 32-channel receiver coil at 3 T, we demonstrate that slice acceleration factors of up to eight (MB=8) with blipped controlled aliasing in parallel imaging (CAIPI), in the absence of in-plane accelerations, can be used routinely with acceptable image quality and integrity for whole brain imaging. Spectral analyses of single-shot fMRI time series demonstrate that temporal fluctuations due to both neuronal and physiological sources were distinguishable and comparable up to slice-acceleration factors of nine (MB=9). The increased temporal efficiency could be employed to achieve, within a given acquisition period, higher spatial resolution, increased fMRI statistical power, multiple TEs, faster sampling of temporal events in a resting state fMRI time series, increased sampling of q-space in diffusion imaging, or more quiet time during a scan.

SUBMITTER: Xu J 

PROVIDER: S-EPMC3815955 | biostudies-literature | 2013 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Evaluation of slice accelerations using multiband echo planar imaging at 3 T.

Xu Junqian J   Moeller Steen S   Auerbach Edward J EJ   Strupp John J   Smith Stephen M SM   Feinberg David A DA   Yacoub Essa E   Uğurbil Kâmil K  

NeuroImage 20130727


We evaluate residual aliasing among simultaneously excited and acquired slices in slice accelerated multiband (MB) echo planar imaging (EPI). No in-plane accelerations were used in order to maximize and evaluate achievable slice acceleration factors at 3 T. We propose a novel leakage (L-) factor to quantify the effects of signal leakage between simultaneously acquired slices. With a standard 32-channel receiver coil at 3 T, we demonstrate that slice acceleration factors of up to eight (MB=8) wit  ...[more]

Similar Datasets

| S-EPMC3429710 | biostudies-literature
| S-EPMC4162886 | biostudies-literature
| S-EPMC6688807 | biostudies-literature
| S-EPMC5923901 | biostudies-other
| S-EPMC6153080 | biostudies-literature
| S-EPMC7166208 | biostudies-literature
| S-EPMC4198428 | biostudies-literature
| S-EPMC7083698 | biostudies-literature
| S-EPMC5448876 | biostudies-other
| S-EPMC5082535 | biostudies-literature