Efficiency of a pneumatic device in controlling cuff pressure of polyurethane-cuffed tracheal tubes: a randomized controlled study.


ABSTRACT: The primary objective of this study was to determine the efficiency of a pneumatic device in controlling cuff pressure (Pcuff) in patients intubated with polyurethane-cuffed tracheal tubes. Secondary objectives were to determine the impact of continuous control of Pcuff, and cuff shape on microaspiration of gastric contents.Prospective randomized controlled study. All patients requiring intubation and mechanical ventilation ?48 h were eligible. The first 32 patients were intubated with tapered polyurethane-cuffed, and the 32 following patients were intubated with cylindrical polyurethane-cuffed tracheal tubes. Patients randomly received 24 h of continuous control of Pcuff using a pneumatic device (Nosten®), and 24 h of routine care of Pcuff using a manometer. Target Pcuff was 25 cmH2O. Pcuff was continuously recorded, and pepsin was quantitatively measured in all tracheal aspirates during these periods.The pneumatic device was efficient in controlling Pcuff (med [IQ] 26 [24, 28] vs 22 [20, 28] cmH2O, during continuous control of Pcuff and routine care, respectively; p?=?0.017). In addition, percentage of patients with underinflation (31% vs 68%) or overinflation (53% vs 100%) of tracheal cuff, and percentage of time spent with underinflation (0.9 [0, 17] vs 14% [4, 30]) or overinflation (0 [0, 2] vs 32% [9, 54]) were significantly (p?

SUBMITTER: Jaillette E 

PROVIDER: S-EPMC3877974 | biostudies-literature | 2013 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Efficiency of a pneumatic device in controlling cuff pressure of polyurethane-cuffed tracheal tubes: a randomized controlled study.

Jaillette Emmanuelle E   Zerimech Farid F   De Jonckheere Julien J   Makris Demosthenes D   Balduyck Malika M   Durocher Alain A   Duhamel Alain A   Nseir Saad S  

BMC anesthesiology 20131226 1


<h4>Background</h4>The primary objective of this study was to determine the efficiency of a pneumatic device in controlling cuff pressure (Pcuff) in patients intubated with polyurethane-cuffed tracheal tubes. Secondary objectives were to determine the impact of continuous control of Pcuff, and cuff shape on microaspiration of gastric contents.<h4>Methods</h4>Prospective randomized controlled study. All patients requiring intubation and mechanical ventilation ≥48 h were eligible. The first 32 pat  ...[more]

Similar Datasets

| S-EPMC5050178 | biostudies-literature
| S-EPMC4461648 | biostudies-literature
| S-EPMC4921025 | biostudies-literature
| S-EPMC5456185 | biostudies-other
| S-EPMC5724199 | biostudies-literature
| S-EPMC7230200 | biostudies-literature
| S-EPMC7642418 | biostudies-literature
| S-EPMC6058997 | biostudies-literature
| S-EPMC7939195 | biostudies-literature
| S-EPMC9276934 | biostudies-literature