Performance comparisons of nano-LC systems, electrospray sources and LC-MS-MS platforms.


ABSTRACT: Selecting a suitable nano-liquid chromatography system (LC), ionization source and mass spectrometer for LC-tandem mass spectrometry (MS-MS) studies is complicated by numerous competing technologies. This study compares four popular nano-LC systems, four ionization sources and three MS facilities that use completely different LC-MS-MS systems. Statistically significant differences in LC performance were identified with similarly performing Proxeon, Waters and Eksigent nanoLC-Ultra systems [retention time routinely at 0.7-0.9% relative standard deviation (RSD)], and all outperformed the Eksigent nanoLC-2D (RSD ?2%). In addition, compatibility issues were identified between the Bruker HCT ion trap mass spectrometer and both the Eksigent nanoLC-2D and the Bruker nanoelectrospray source. The electrospray source itself had an unexpected and striking effect on chromatographic reproducibility on the Bruker HCT ion trap. The New Objective nanospray source significantly outperformed the Bruker nanospray source in retention time RSD (1% RSD versus 14% RSD, respectively); and the Bruker nebulized nanospray source outperformed both of these traditional, non-nebulized sources (0.5% RSD in retention time). Finally, to provide useful benchmarks for overall proteomics sensitivity, different LC-MS-MS platforms were compared by analyzing a range of concentrations of tryptic digests of bovine serum albumin at three MS facilities. The results indicate that similar sensitivity can be realized with a Bruker HCT-Ultra ion trap, a Thermo LTQ-Velos Linear ion trap and a Thermo LTQ-Orbitrap XL-ETD.

SUBMITTER: Liu Q 

PROVIDER: S-EPMC3886563 | biostudies-literature | 2014 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance comparisons of nano-LC systems, electrospray sources and LC-MS-MS platforms.

Liu Qian Q   Cobb Jennifer S JS   Johnson Joshua L JL   Wang Qi Q   Agar Jeffrey N JN  

Journal of chromatographic science 20130117 2


Selecting a suitable nano-liquid chromatography system (LC), ionization source and mass spectrometer for LC-tandem mass spectrometry (MS-MS) studies is complicated by numerous competing technologies. This study compares four popular nano-LC systems, four ionization sources and three MS facilities that use completely different LC-MS-MS systems. Statistically significant differences in LC performance were identified with similarly performing Proxeon, Waters and Eksigent nanoLC-Ultra systems [reten  ...[more]

Similar Datasets

| S-EPMC3319463 | biostudies-literature
| S-EPMC3901191 | biostudies-literature
| S-EPMC4113806 | biostudies-literature
| S-EPMC3730131 | biostudies-other
| S-EPMC8374884 | biostudies-literature
| S-EPMC10408068 | biostudies-literature
| S-EPMC5606153 | biostudies-literature
| S-EPMC2830837 | biostudies-other
| S-EPMC4062842 | biostudies-literature
| S-EPMC4684172 | biostudies-literature