Unknown

Dataset Information

0

Microbiopsy engineered for minimally invasive and suture-free sub-millimetre skin sampling.


ABSTRACT: We describe the development of a sub-millimetre skin punch biopsy device for minimally invasive and suture-free skin sampling for molecular diagnosis and research. Conventional skin punch biopsies range from 2-4 mm in diameter. Local anaesthesia is required and sutures are usually used to close the wound. Our microbiopsy is 0.50 mm wide and 0.20 mm thick. The microbiopsy device is fabricated from three stacked medical grade stainless steel plates tapered to a point and contains a chamber within the centre plate to collect the skin sample. We observed that the application of this device resulted in a 0.21 ± 0.04 mm wide puncture site in volunteer skin using reflectance confocal microscopy. Histological sections from microbiopsied skin revealed 0.22 ± 0.12 mm wide and 0.26 ± 0.09 mm deep puncture sites. Longitudinal observation in microbiopsied volunteers showed that the wound closed within 1 day and was not visible after 7 days. Reflectance confocal microscope images from these same sites showed the formation of a tiny crust that resolved by 3 weeks and was completely undetectable by the naked eye. The design parameters of the device were optimised for molecular analysis using sampled DNA mass as the primary end point in volunteer studies. Finally, total RNA was characterized. The optimised device extracted 5.9 ± 3.4 ng DNA and 9.0 ± 10.1 ng RNA. We foresee that minimally invasive molecular sampling will play an increasingly significant role in diagnostic dermatology and skin research.

SUBMITTER: Lin LL 

PROVIDER: S-EPMC3907159 | biostudies-literature | 2013

REPOSITORIES: biostudies-literature

altmetric image

Publications

Microbiopsy engineered for minimally invasive and suture-free sub-millimetre skin sampling.

Lin Lynlee L LL   Prow Tarl W TW   Raphael Anthony P AP   Harrold Iii Robert L RL   Primiero Clare A CA   Ansaldo Alexander B AB   Soyer H Peter HP  

F1000Research 20130502


We describe the development of a sub-millimetre skin punch biopsy device for minimally invasive and suture-free skin sampling for molecular diagnosis and research. Conventional skin punch biopsies range from 2-4 mm in diameter. Local anaesthesia is required and sutures are usually used to close the wound. Our microbiopsy is 0.50 mm wide and 0.20 mm thick. The microbiopsy device is fabricated from three stacked medical grade stainless steel plates tapered to a point and contains a chamber within  ...[more]

Similar Datasets

| S-EPMC9478331 | biostudies-literature
| S-EPMC11323771 | biostudies-literature
| S-EPMC7460368 | biostudies-literature
| S-EPMC6746174 | biostudies-literature
| S-EPMC6897551 | biostudies-literature
| S-EPMC8672862 | biostudies-literature
2022-04-06 | PXD029003 | Pride
| S-EPMC6706129 | biostudies-literature
| S-EPMC7839767 | biostudies-literature