Unknown

Dataset Information

0

A multiplex two-color real-time PCR method for quality-controlled molecular diagnostic testing of FFPE samples.


ABSTRACT:

Background

Reverse transcription quantitative real-time PCR (RT-qPCR) tests support personalized cancer treatment through more clinically meaningful diagnosis. However, samples obtained through standard clinical pathology procedures are formalin-fixed, paraffin-embedded (FFPE) and yield small samples with low integrity RNA containing PCR interfering substances. RT-qPCR tests able to assess FFPE samples with quality control and inter-laboratory reproducibility are needed.

Methods

We developed an RT-qPCR method by which 1) each gene was measured relative to a known number of its respective competitive internal standard molecules to control for interfering substances, 2) two-color fluorometric hydrolysis probes enabled analysis on a real-time platform, 3) external standards controlled for variation in probe fluorescence intensity, and 4) pre-amplification maximized signal from FFPE RNA samples. Reagents were developed for four genes comprised by a previously reported lung cancer diagnostic test (LCDT) then subjected to analytical validation using synthetic native templates as test articles to assess linearity, signal-to-analyte response, lower detection threshold, imprecision and accuracy. Fitness of this method and these reagents for clinical testing was assessed in FFPE normal (N = 10) and malignant (N = 10) lung samples.

Results

Reagents for each of four genes, MYC, E2F1, CDKN1A and ACTB comprised by the LCDT had acceptable linearity (R(2)>0.99), signal-to-analyte response (slope 1.0 ± 0.05), lower detection threshold (<10 molecules) and imprecision (CV <20%). Poisson analysis confirmed accuracy of internal standard concentrations. Internal standards controlled for experimentally introduced interference, prevented false-negatives and enabled pre-amplification to increase signal without altering measured values. In the fitness for purpose testing of this two-color fluorometric LCDT using surgical FFPE samples, the diagnostic accuracy was 93% which was similar to that previously reported for analysis of fresh samples.

Conclusions

This quality-controlled two-color fluorometric RT-qPCR approach will facilitate the development of reliable, robust RT-qPCR-based molecular diagnostic tests in FFPE clinical samples.

SUBMITTER: Yeo J 

PROVIDER: S-EPMC3931751 | biostudies-literature | 2014

REPOSITORIES: biostudies-literature

altmetric image

Publications

A multiplex two-color real-time PCR method for quality-controlled molecular diagnostic testing of FFPE samples.

Yeo Jiyoun J   Crawford Erin L EL   Blomquist Thomas M TM   Stanoszek Lauren M LM   Dannemiller Rachel E RE   Zyrek Jill J   De Las Casas Luis E LE   Khuder Sadik A SA   Willey James C JC  

PloS one 20140221 2


<h4>Background</h4>Reverse transcription quantitative real-time PCR (RT-qPCR) tests support personalized cancer treatment through more clinically meaningful diagnosis. However, samples obtained through standard clinical pathology procedures are formalin-fixed, paraffin-embedded (FFPE) and yield small samples with low integrity RNA containing PCR interfering substances. RT-qPCR tests able to assess FFPE samples with quality control and inter-laboratory reproducibility are needed.<h4>Methods</h4>W  ...[more]

Similar Datasets

| S-EPMC3561136 | biostudies-literature
| S-EPMC5494621 | biostudies-literature
| S-EPMC2897455 | biostudies-literature
| S-EPMC4786821 | biostudies-other
| S-EPMC3536235 | biostudies-literature