Comparison of the illumigene Mycoplasma DNA amplification assay and culture for detection of Mycoplasma pneumoniae.


ABSTRACT: A loop-mediated isothermal amplification (LAMP) system, the illumigene Mycoplasma DNA amplification assay (Meridian Bioscience, Inc., Cincinnati, OH) was evaluated to determine its analytical sensitivity, specificity, and clinical application in comparison to historic culture in a collection of archived respiratory specimens. The illumigene limit of detection was ?88 CFU/reaction for 10 Mycoplasma pneumoniae reference strains. This assay correctly identified 36 M. pneumoniae reference strains and clinical isolates from various geographic origins, including both of the main subtypes. No cross-reactions were detected with other mycoplasmas, ureaplasmas, other bacterial species, viruses, yeasts, or human DNA. Among 214 respiratory specimens previously cultured for M. pneumoniae, when real-time PCR with bidirectional sequencing of the PCR products was used to resolve discrepancies, the sensitivity was 22 of 22 (100%) and the specificity was 190 of 192 (99%). This commercial LAMP assay is a useful rapid method for detecting M. pneumoniae in clinical specimens. Additional prospective clinical trials with direct comparison to culture and PCR are warranted.

SUBMITTER: Ratliff AE 

PROVIDER: S-EPMC3993520 | biostudies-literature | 2014 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of the illumigene Mycoplasma DNA amplification assay and culture for detection of Mycoplasma pneumoniae.

Ratliff Amy E AE   Duffy Lynn B LB   Waites Ken B KB  

Journal of clinical microbiology 20140115 4


A loop-mediated isothermal amplification (LAMP) system, the illumigene Mycoplasma DNA amplification assay (Meridian Bioscience, Inc., Cincinnati, OH) was evaluated to determine its analytical sensitivity, specificity, and clinical application in comparison to historic culture in a collection of archived respiratory specimens. The illumigene limit of detection was ≤88 CFU/reaction for 10 Mycoplasma pneumoniae reference strains. This assay correctly identified 36 M. pneumoniae reference strains an  ...[more]

Similar Datasets

| S-EPMC6023700 | biostudies-literature
| S-EPMC6988361 | biostudies-literature
| S-EPMC3341502 | biostudies-literature
| S-EPMC9000086 | biostudies-literature
| S-EPMC4625007 | biostudies-literature
| S-EPMC6275423 | biostudies-literature
| S-EPMC5527399 | biostudies-literature
| S-EPMC6717426 | biostudies-literature
| S-EPMC4540935 | biostudies-literature
| S-EPMC6767991 | biostudies-literature