ABSTRACT: Introduction
Mobile phones hold considerable promise for delivering evidence-based smoking cessation interventions that require frequent and objective assessment of smoking status via breath carbon monoxide (Breath CO) measurement. However, there are currently no commercially available mobile-phone-based Breath CO meters. We developed a mobile-phone-based Breath CO meter prototype that attaches to and communicates with a smartphone through an audio port. We then evaluated the reliability and the validity of Breath CO measures collected with the mobile meter prototype and assessed the usability and acceptability of the meter.Methods
Participants included 20 regular smokers (?10 cigarettes/day), 20 light smokers (<10 cigarettes/day), and 20 nonsmokers. Expired air samples were collected 4 times from each participant: twice with the mobile meter and twice with a commercially available Breath CO meter.Results
Measures calculated by the mobile meter correlated strongly with measures calculated by the commercial meter (r = .96, p < .001). Additionally, the mobile meter accurately distinguished between smokers and nonsmokers. The area under the receiver-operating characteristic curve for the mobile meter was 94.7%, and the meter had a combined sensitivity and specificity of 1.86 at an abstinence threshold of ?6 ppm. Responses on an acceptability survey indicated that smokers liked the meter and would be interested in using it during a quit attempt.Conclusions
The results of our study suggest that a mobile-phone-based Breath CO meter is a reliable, valid, and acceptable device for distinguishing between smokers and nonsmokers.
SUBMITTER: Meredith SE
PROVIDER: S-EPMC4031569 | biostudies-literature | 2014 Jun
REPOSITORIES: biostudies-literature
Meredith Steven E SE Robinson Andrew A Erb Philip P Spieler Claire A CA Klugman Noah N Dutta Prabal P Dallery Jesse J
Nicotine & tobacco research : official journal of the Society for Research on Nicotine and Tobacco 20140127 6
<h4>Introduction</h4>Mobile phones hold considerable promise for delivering evidence-based smoking cessation interventions that require frequent and objective assessment of smoking status via breath carbon monoxide (Breath CO) measurement. However, there are currently no commercially available mobile-phone-based Breath CO meters. We developed a mobile-phone-based Breath CO meter prototype that attaches to and communicates with a smartphone through an audio port. We then evaluated the reliability ...[more]