Validation of the only commercially available immunoassay for synthetic cathinones in urine: Randox Drugs of Abuse V Biochip Array Technology.


ABSTRACT: Deterrence of synthetic cathinone abuse is hampered by the lack of a high-throughput immunoassay screen. The Randox Drugs of Abuse V (DOA-V) Biochip Array Technology contains two synthetic cathinone antibodies: Bath Salt I (BSI) targets mephedrone/methcathinone and Bath Salt II (BSII) targets 3',4'-methylenedioxypyrovalerone (MDPV)/3',4'-methylenedioxy-?-pyrrolidinobutiophenone (MDPBP). We evaluated DOA-V synthetic cathinones performance and conducted a full validation on the original assay with calibrators reconstituted in water, and the new assay with calibrators prepared in lyophilized urine; both utilized the same antibodies and were run on the fully automated Evidence® Analyzer. We screened 20 017 authentic military urine specimens and confirmed positives by liquid chromatography-tandem mass spectrometry (LC-MS/MS) for 28 synthetic cathinones. Limits of detection (LOD) for the original and new assays were 0.35 and 0.18 (BSI), and 8.5 and 9.2 µg/L (BSII), respectively. Linearity was acceptable (R(2) ?>0.98); however, a large negative bias was observed with in-house prepared calibrators. Intra-assay imprecision was <20% BSI-II, while inter-assay imprecision was 18-42% BSI and <22% BSII. Precision was acceptable for Randox controls. Cross-reactivities of many additional synthetic cathinones were determined. Authentic drug-free negative urine pH <4 produced false positive results for BSI (6.3 µg/L) and BSII (473 µg/L). Oxidizing agents reduced BSI and increased BSII results. Sensitivity, specificity, and efficiency of 100%, 52.1%, and 53.0% were obtained at manufacturer's proposed cut-offs (BSI 5 µg/L, BSII 30 µg/L). Performance improved if cut-off concentrations increased (BSI 7.5 µg/L, BSII 40 µg/L); however, there were limited confirmed positive specimens. Currently, this is the first and only fully validated immunoassay for preliminary detection of synthetic cathinones in urine. Published 2014. This article is a U.S. Government work and is in the public domain in the USA.

SUBMITTER: Ellefsen KN 

PROVIDER: S-EPMC4107059 | biostudies-literature | 2014 Jul-Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Validation of the only commercially available immunoassay for synthetic cathinones in urine: Randox Drugs of Abuse V Biochip Array Technology.

Ellefsen Kayla N KN   Anizan Sébastien S   Castaneto Marisol S MS   Desrosiers Nathalie A NA   Martin Thomas M TM   Klette Kevin L KL   Huestis Marilyn A MA  

Drug testing and analysis 20140321 7-8


Deterrence of synthetic cathinone abuse is hampered by the lack of a high-throughput immunoassay screen. The Randox Drugs of Abuse V (DOA-V) Biochip Array Technology contains two synthetic cathinone antibodies: Bath Salt I (BSI) targets mephedrone/methcathinone and Bath Salt II (BSII) targets 3',4'-methylenedioxypyrovalerone (MDPV)/3',4'-methylenedioxy-α-pyrrolidinobutiophenone (MDPBP). We evaluated DOA-V synthetic cathinones performance and conducted a full validation on the original assay with  ...[more]

Similar Datasets

| S-EPMC7320699 | biostudies-literature
| S-EPMC5556313 | biostudies-other
| S-EPMC7080616 | biostudies-literature
| S-EPMC5340536 | biostudies-literature
| S-EPMC3402036 | biostudies-literature
2021-08-01 | GSE156973 | GEO
| S-EPMC8025544 | biostudies-literature
| S-EPMC4834613 | biostudies-other
2019-09-27 | GSE133719 | GEO
| S-EPMC6479684 | biostudies-literature