Experimental evaluation and optimization of structures for lossless ion manipulations for ion mobility spectrometry with time-of-flight mass spectrometry.


ABSTRACT: We report on the performance of structures for lossless ion manipulation (SLIM) as a means for transmitting ions and performing ion mobility separations (IMS). Ions were successfully transferred from an electrospray ionization (ESI) source to the TOF MS analyzer by means of a linear SLIM, demonstrating lossless ion transmission and an alternative arrangement including a 90° turn. First, the linear geometry was optimized for radial confinement by tuning RF on the central "rung" electrodes and potentials on the DC-only guard electrodes. Selecting an appropriate DC guard bias (2-6 V) and RF amplitude (?160 V(p-p) at 750 kHz) resulted in the greatest ion intensities. Close to ideal IMS resolving power was maintained over a significant range of applied voltages. Second, the 90° turn was optimized for radial confinement by tuning RF on the rung electrodes and DC on the guard electrodes. However, both resolving power and ion transmission showed a dependence on these voltages, and the best conditions for both were >300 V(p-p) RF (685 kHz) and 7-11 V guard DC bias. Both geometries provide IMS resolving powers at the theoretical limit (R ~ 58), showing that degraded resolution from a "racetrack" effect from turning around a corner can be successfully avoided, and the capability also was maintained for essentially lossless ion transmission.

SUBMITTER: Webb IK 

PROVIDER: S-EPMC4165449 | biostudies-literature | 2014 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Experimental evaluation and optimization of structures for lossless ion manipulations for ion mobility spectrometry with time-of-flight mass spectrometry.

Webb Ian K IK   Garimella Sandilya V B SV   Tolmachev Aleksey V AV   Chen Tsung-Chi TC   Zhang Xinyu X   Norheim Randolph V RV   Prost Spencer A SA   LaMarche Brian B   Anderson Gordon A GA   Ibrahim Yehia M YM   Smith Richard D RD  

Analytical chemistry 20140905 18


We report on the performance of structures for lossless ion manipulation (SLIM) as a means for transmitting ions and performing ion mobility separations (IMS). Ions were successfully transferred from an electrospray ionization (ESI) source to the TOF MS analyzer by means of a linear SLIM, demonstrating lossless ion transmission and an alternative arrangement including a 90° turn. First, the linear geometry was optimized for radial confinement by tuning RF on the central "rung" electrodes and pot  ...[more]

Similar Datasets

| S-EPMC4188271 | biostudies-other
| S-EPMC5603265 | biostudies-literature
| S-EPMC6211290 | biostudies-literature
| S-EPMC4793926 | biostudies-literature
| S-EPMC4748841 | biostudies-literature
| S-EPMC5627995 | biostudies-literature
| S-EPMC8216491 | biostudies-literature
| S-EPMC9048586 | biostudies-literature
| S-EPMC5815952 | biostudies-literature
| S-EPMC6191335 | biostudies-literature