Ontology highlight
ABSTRACT: Introduction
Spirometry should follow strict quality criteria. The American Thoracic Society (ATS) recommends the use of a noseclip; however there are controversies about its need. ATS also indicates that tests should be done in the sitting position, but there are no recommendations neither about position of the upper limbs and lower limbs nor about who should hold the mouthpiece while performing the maneuvers: evaluated subject or evaluator.Objectives
To compare noseclip use or not, different upper and lower limbs positions and who holds the mouthpiece, verifying if these technical details affect spirometric results in healthy adults.Methods
One hundred and three healthy individuals (41 men; age: 47 [33-58] years; normal lung function: FEV?/FVC?=?83±5, FEV??=?94 [88-104]%predicted, FVC?=?92 [84-102]%predicted) underwent a protocol consisting of four spirometric comparative analysis in the sitting position: 1) maximum voluntary ventilation (MVV) with vs without noseclip; 2) FVC performed with vs without upper limbs support; 3) FVC performed with lower limbs crossed vs lower limbs in neutral position; 4) FVC, slow vital capacity and MVV comparing the evaluated subject holding the mouthpiece vs evaluator holding it.Results
Different spirometric variables presented statistically significant difference (p<0.05) when analysing the four comparisons; however, none of them showed any variation larger than those considered as acceptable according to the ATS reproducibility criteria.Conclusions
There was no relevant variation in spirometric results when analyzing technical details such as noseclip use during MVV, upper and lower limb positions and who holds the mouthpiece when performing the tests in healthy adults.
SUBMITTER: Sipoli L
PROVIDER: S-EPMC4171496 | biostudies-literature | 2014
REPOSITORIES: biostudies-literature
Sipoli Luciana L Martinez Larissa L Donária Leila L Probst Vanessa Suziane VS Moreira Graciane Laender GL Pitta Fabio F
PloS one 20140922 9
<h4>Introduction</h4>Spirometry should follow strict quality criteria. The American Thoracic Society (ATS) recommends the use of a noseclip; however there are controversies about its need. ATS also indicates that tests should be done in the sitting position, but there are no recommendations neither about position of the upper limbs and lower limbs nor about who should hold the mouthpiece while performing the maneuvers: evaluated subject or evaluator.<h4>Objectives</h4>To compare noseclip use or ...[more]