Performance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics E170 Total 25-OH Vitamin D assays.
Ontology highlight
ABSTRACT: OBJECTIVES:To evaluate the newly developed Roche MODULAR Analytics E170 Total Vitamin D and the Siemens ADVIA Centaur Vitamin D Total assays. MATERIALS AND METHODS:Assays were evaluated using the Clinical and Laboratory Standards Institute protocols. Split patient samples were compared with LC-MS/MS and DiaSorin LIAISON assays (n=79 including 15 specimens with detectable endogenous 25-OH vitamin D(2)). Assay accuracy was also evaluated using the Vitamin D External Quality Assessment Scheme (DEQAS) samples. RESULTS:The ADVIA Centaur and E170 assays demonstrated maximum total CVs of 14.1% and 5.9%, respectively. Both showed excellent linearity (R(2)>0.99). The ADVIA Centaur assay demonstrated interference with bilirubin at 800 ?mol/L, hemolysis at 1.25 g/L, and triglycerides at 2.8 mmol/L. Compared to LC-MS/MS, the ADVIA Centaur assay demonstrated a R(2) value of 0.893, average bias of -8.8%; the E170 assay an R(2) value of 0.872, average bias of 14.3% with underestimation of 25-OH vitamin D(2). Compared to the LIAISON assay, the ADVIA Centaur assay demonstrated an R(2) value of 0.781, average bias of -17.3%; the E170 assay an R(2) value of 0.823, average bias of 11.4%. The ADVIA Centaur and E170 assays demonstrated a biases of <20% in 10/10 and 8/10 DEQAS samples, respectively. CONCLUSIONS:The ADVIA Centaur and E170 vitamin D assays demonstrated acceptable linearity, imprecision, and accuracy. The E170 assay demonstrated consistent underestimation of 25-OH vitamin D(2) levels. Compared with LC-MS/MS, the ADVIA Centaur assay demonstrated a higher R(2) value and a smaller average bias than the E170 assay.
SUBMITTER: Chen Y
PROVIDER: S-EPMC4193681 | biostudies-literature | 2012 Nov
REPOSITORIES: biostudies-literature
ACCESS DATA