Unknown

Dataset Information

0

Performance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics E170 Total 25-OH Vitamin D assays.


ABSTRACT: OBJECTIVES:To evaluate the newly developed Roche MODULAR Analytics E170 Total Vitamin D and the Siemens ADVIA Centaur Vitamin D Total assays. MATERIALS AND METHODS:Assays were evaluated using the Clinical and Laboratory Standards Institute protocols. Split patient samples were compared with LC-MS/MS and DiaSorin LIAISON assays (n=79 including 15 specimens with detectable endogenous 25-OH vitamin D(2)). Assay accuracy was also evaluated using the Vitamin D External Quality Assessment Scheme (DEQAS) samples. RESULTS:The ADVIA Centaur and E170 assays demonstrated maximum total CVs of 14.1% and 5.9%, respectively. Both showed excellent linearity (R(2)>0.99). The ADVIA Centaur assay demonstrated interference with bilirubin at 800 ?mol/L, hemolysis at 1.25 g/L, and triglycerides at 2.8 mmol/L. Compared to LC-MS/MS, the ADVIA Centaur assay demonstrated a R(2) value of 0.893, average bias of -8.8%; the E170 assay an R(2) value of 0.872, average bias of 14.3% with underestimation of 25-OH vitamin D(2). Compared to the LIAISON assay, the ADVIA Centaur assay demonstrated an R(2) value of 0.781, average bias of -17.3%; the E170 assay an R(2) value of 0.823, average bias of 11.4%. The ADVIA Centaur and E170 assays demonstrated a biases of <20% in 10/10 and 8/10 DEQAS samples, respectively. CONCLUSIONS:The ADVIA Centaur and E170 vitamin D assays demonstrated acceptable linearity, imprecision, and accuracy. The E170 assay demonstrated consistent underestimation of 25-OH vitamin D(2) levels. Compared with LC-MS/MS, the ADVIA Centaur assay demonstrated a higher R(2) value and a smaller average bias than the E170 assay.

SUBMITTER: Chen Y 

PROVIDER: S-EPMC4193681 | biostudies-literature | 2012 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Performance evaluation of Siemens ADVIA Centaur and Roche MODULAR Analytics E170 Total 25-OH Vitamin D assays.

Chen Yu Y   Kinney Lois L   Božović Andrea A   Smith Hilary H   Tarr Heather H   Diamandis Eleftherios P EP   LeBlanc Adrien A  

Clinical biochemistry 20120613 16-17


<h4>Objectives</h4>To evaluate the newly developed Roche MODULAR Analytics E170 Total Vitamin D and the Siemens ADVIA Centaur Vitamin D Total assays.<h4>Materials and methods</h4>Assays were evaluated using the Clinical and Laboratory Standards Institute protocols. Split patient samples were compared with LC-MS/MS and DiaSorin LIAISON assays (n=79 including 15 specimens with detectable endogenous 25-OH vitamin D(2)). Assay accuracy was also evaluated using the Vitamin D External Quality Assessme  ...[more]

Similar Datasets

| S-EPMC4567539 | biostudies-literature
| S-EPMC4763092 | biostudies-literature
| S-EPMC5104342 | biostudies-literature
| S-EPMC5121522 | biostudies-literature
| S-EPMC6807057 | biostudies-literature
| S-EPMC6589729 | biostudies-literature
| 2110440 | ecrin-mdr-crc
| S-EPMC4012279 | biostudies-other
| S-EPMC6126881 | biostudies-literature
| S-EPMC6695248 | biostudies-literature