AQP4 autoantibody assay performance in clinical laboratory service.


ABSTRACT: To compare performance of contemporary aquaporin-4-immunoglobulin (Ig) G assays in clinical service.Sera from neurologic patients (4 groups) and controls were tested initially by service ELISA (recombinant human aquaporin-4, M1 isoform) and then by cell-based fluorescence assays: fixed (CBA, M1-aquaporin-4, observer-scored) and live (fluorescence-activated cell sorting [FACS], M1 and M23 aquaporin-4 isoforms). Group 1: all Mayo Clinic patients tested from January to May 2012; group 2: consecutive aquaporin-4-IgG-positive patients from September 2011 (Mayo and non-Mayo); group 3: suspected ELISA false-negatives from 2011 to 2013 (physician-reported, high likelihood of neuromyelitis optica spectrum disorders [NMOSDs] clinically); group 4: suspected ELISA false-positives (physician-reported, not NMOSD clinically).Group 1 (n = 388): M1-FACS assay performed optimally (areas under the curves: M1 = 0.64; M23 = 0.57 [p = 0.02]). Group 2 (n = 30): NMOSD clinical diagnosis was confirmed by: M23-FACS, 24; M1-FACS, 23; M1-CBA, 20; and M1-ELISA, 18. Six results were suspected false-positive: M23-FACS, 2; M1-ELISA, 2; and M23-FACS, M1-FACS, and M1-CBA, 2. Group 3 (n = 31, suspected M1-ELISA false-negatives): results were positive for 5 sera: M1-FACS, 5; M23-FACS, 3; and M1-CBA, 2. Group 4 (n = 41, suspected M1-ELISA false-positives): all negative except 1 (positive only by M1-CBA). M1/M23-cotransfected cells expressing smaller membrane arrays of aquaporin-4 yielded fewer false- positive FACS results than M23-transfected cells.Aquaporin-4-transfected CBAs, particularly M1-FACS, perform optimally in aiding NMOSD serologic diagnosis. High-order arrays of M23-aquaporin-4 may yield false-positive results by binding IgG nonspecifically.

SUBMITTER: Fryer JP 

PROVIDER: S-EPMC4202686 | biostudies-literature | 2014 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

AQP4 autoantibody assay performance in clinical laboratory service.

Fryer J P JP   Lennon V A VA   Pittock S J SJ   Jenkins S M SM   Fallier-Becker P P   Clardy S L SL   Horta E E   Jedynak E A EA   Lucchinetti C F CF   Shuster E A EA   Weinshenker B G BG   Wingerchuk D M DM   McKeon A A  

Neurology(R) neuroimmunology & neuroinflammation 20140522 1


<h4>Objective</h4>To compare performance of contemporary aquaporin-4-immunoglobulin (Ig) G assays in clinical service.<h4>Methods</h4>Sera from neurologic patients (4 groups) and controls were tested initially by service ELISA (recombinant human aquaporin-4, M1 isoform) and then by cell-based fluorescence assays: fixed (CBA, M1-aquaporin-4, observer-scored) and live (fluorescence-activated cell sorting [FACS], M1 and M23 aquaporin-4 isoforms). Group 1: all Mayo Clinic patients tested from Januar  ...[more]

Similar Datasets

| S-EPMC8936135 | biostudies-literature
| S-EPMC10036445 | biostudies-literature
| S-EPMC8631108 | biostudies-literature
| S-EPMC5497918 | biostudies-literature
| S-EPMC6536857 | biostudies-literature
| S-EPMC10481325 | biostudies-literature
| S-EPMC5033450 | biostudies-other
2019-02-25 | GSE99573 | GEO
| S-EPMC5598812 | biostudies-literature
| S-EPMC6499319 | biostudies-literature