Randomised masked clinical trial of the MGDRx EyeBag for the treatment of meibomian gland dysfunction-related evaporative dry eye.


ABSTRACT: To investigate the efficacy and safety of the MGDRx EyeBag (The Eyebag Company, Halifax, UK) eyelid warming device.Twenty-five patients with confirmed meibomian gland dysfunction (MGD)-related evaporative dry eye were enrolled into a randomised, single masked, contralateral clinical trial. Test eyes received a heated device; control eyes a non-heated device for 5?min twice a day for 2?weeks. Efficacy (ocular symptomology, non-invasive break-up time, lipid layer thickness, osmolarity, meibomian gland dropout and function) and safety (visual acuity, corneal topography, conjunctival hyperaemia and staining) measurements were taken at baseline and follow-up. Subsequent patient device usage and ocular comfort was ascertained at 6?months.Differences between test and control eyes at baseline were not statistically significant for all measurements (p>0.05). After 2?weeks, statistically significant improvements occurred in all efficacy measurements in test eyes (p<0.05). Visual acuity and corneal topography were unaffected (p>0.05). All patients maintained higher ocular comfort after 6?months (p<0.05), although the benefit was greater in those who continued usage 1-8 times a month (p<0.001).The MGDRx EyeBag is a safe and effective device for the treatment of MGD-related evaporative dry eye. Subjective benefit lasts at least 6?months, aided by occasional retreatment.NCT01870180.

SUBMITTER: Bilkhu PS 

PROVIDER: S-EPMC4251196 | biostudies-literature | 2014 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Randomised masked clinical trial of the MGDRx EyeBag for the treatment of meibomian gland dysfunction-related evaporative dry eye.

Bilkhu Paramdeep Singh PS   Naroo Shehzad Anjam SA   Wolffsohn James Stuart JS  

The British journal of ophthalmology 20140704 12


<h4>Background/aims</h4>To investigate the efficacy and safety of the MGDRx EyeBag (The Eyebag Company, Halifax, UK) eyelid warming device.<h4>Methods</h4>Twenty-five patients with confirmed meibomian gland dysfunction (MGD)-related evaporative dry eye were enrolled into a randomised, single masked, contralateral clinical trial. Test eyes received a heated device; control eyes a non-heated device for 5 min twice a day for 2 weeks. Efficacy (ocular symptomology, non-invasive break-up time, lipid  ...[more]

Similar Datasets

| S-EPMC5685175 | biostudies-other
| S-EPMC4968866 | biostudies-literature
| S-EPMC10959456 | biostudies-literature
| S-EPMC8564894 | biostudies-literature
| S-EPMC7763017 | biostudies-literature
| S-EPMC8673856 | biostudies-literature
| S-EPMC10323053 | biostudies-literature
| S-EPMC3770344 | biostudies-literature
| S-EPMC8719179 | biostudies-literature
| S-EPMC9109882 | biostudies-literature