MICROFLUIDIC DEVICES FOR LABEL-FREE AND NON-INSTRUMENTED QUANTITATION OF UNAMPLIFIED NUCLEIC ACIDS BY FLOW DISTANCE MEASUREMENT.


ABSTRACT: Timely biomarker quantitation has potential to improve human health but current methods have disadvantages either in terms of cost and complexity for benchtop instruments, or reduced performance in quantitation and/or multiplexing for point-of-care systems. We previously developed microfluidic devices wherein visually observed flow distances correlated with a model analyte's concentration.1 Here, we significantly expand over this prior result to demonstrate the measurement of unamplified DNA analogues of microRNAs (miRNAs), biomarkers whose levels can be altered in disease states. We have developed a method for covalently attaching nucleic acid receptors on poly(dimethylsiloxane) microchannel surfaces by silane and cross-linker treatments. We found a flow distance dependence on target concentrations from 10 ?g/mL to 10 pg/mL for DNA in both buffer and synthetic urine. Moreover, flow time in addition to flow distance is correlated with target concentration. We also observed longer flow distances for single-base mismatches compared to the target sequence at the same concentration, indicating that our approach can be used to detect point mutations. Finally, experiments with DNA analogues of miRNA biomarkers for kidney disease (mir-200c-3p) and prostate cancer (mir-107) in synthetic urine showed the ability to detect these analytes near clinically relevant levels. Our results demonstrate that these novel microfluidic assays offer a simple route to sensitive, amplification-free nucleic acid quantitation, with strong potential for point-of-care application.

SUBMITTER: Chatterjee D 

PROVIDER: S-EPMC4269297 | biostudies-literature | 2014 Oct

REPOSITORIES: biostudies-literature

altmetric image

Publications

MICROFLUIDIC DEVICES FOR LABEL-FREE AND NON-INSTRUMENTED QUANTITATION OF UNAMPLIFIED NUCLEIC ACIDS BY FLOW DISTANCE MEASUREMENT.

Chatterjee Debolina D   Mansfield Danielle S DS   Woolley Adam T AT  

Analytical methods : advancing methods and applications 20141001 20


Timely biomarker quantitation has potential to improve human health but current methods have disadvantages either in terms of cost and complexity for benchtop instruments, or reduced performance in quantitation and/or multiplexing for point-of-care systems. We previously developed microfluidic devices wherein visually observed flow distances correlated with a model analyte's concentration.<sup>1</sup> Here, we significantly expand over this prior result to demonstrate the measurement of unamplif  ...[more]

Similar Datasets

| S-EPMC5890436 | biostudies-literature
| S-EPMC5070811 | biostudies-literature
| S-EPMC9035142 | biostudies-literature
| S-EPMC5473339 | biostudies-literature
| S-EPMC3273337 | biostudies-literature
| S-EPMC3365910 | biostudies-other
| S-EPMC8125272 | biostudies-literature
| S-EPMC3558029 | biostudies-literature
| S-EPMC5321777 | biostudies-literature
| S-EPMC3516876 | biostudies-literature