Unknown

Dataset Information

Routine serum creatinine measurements: how well do we perform?


ABSTRACT:

Background

The first aim of the study was to investigate the accuracy and intra-laboratory variation of serum creatinine measurements in clinical laboratories in Flanders. The second purpose was to check the effect of this variation in serum creatinine concentration results on the calculated estimated glomerular filtration rate (eGFR) and the impact on classification of patients into a chronic kidney disease (CKD) stage.

Methods

26 routine instruments were included, representing 13 different types of analyzers from 6 manufacturers and covering all current methodologies (Jaffe, compensated Jaffe, enzymatic liquid and dry chemistry methods). Target values of five serum pools (creatinine concentrations ranging from 35 to 934 μmol/L) were assigned by the gold standard method (ID

SUBMITTER: Hoste L 