Quantitative laser diffraction method for the assessment of protein subvisible particles.


ABSTRACT: Laser diffraction (LD) has been recognized as a method for estimating particle size distribution. Here, a recently developed quantitative LD (qLD) system, which is an LD method with extensive deconvolution analysis, was employed for the quantitative assessment of protein particles sizes, especially aimed at the quantification of 0.2-10 ?m diameter subvisible particles (SVPs). The qLD accurately estimated concentration distributions for silica beads with diameters ranging from 0.2 to 10 ?m that have refractive indices similar to that of protein particles. The linearity of concentration for micrometer-diameter silica beads was confirmed in the presence of a fixed concentration of submicrometer diameter beads. Similarly, submicrometer-diameter silica beads could be quantified in the presence of micrometer-diameter beads. Subsequently, stir- and heat-stressed intravenous immunoglobulins were evaluated by using the qLD, in which the refractive index of protein particles that was determined experimentally was used in the deconvolution analysis. The results showed that the concentration distributions of protein particles in SVP size range differ for the two stresses. The number concentration of the protein particles estimated using the qLD agreed well with that obtained using flow microscopy. This work demonstrates that qLD can be used for quantitative estimation of protein aggregates in SVP size range.

SUBMITTER: Totoki S 

PROVIDER: S-EPMC4359019 | biostudies-literature | 2015 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Quantitative laser diffraction method for the assessment of protein subvisible particles.

Totoki Shinichiro S   Yamamoto Gaku G   Tsumoto Kouhei K   Uchiyama Susumu S   Fukui Kiichi K  

Journal of pharmaceutical sciences 20141201 2


Laser diffraction (LD) has been recognized as a method for estimating particle size distribution. Here, a recently developed quantitative LD (qLD) system, which is an LD method with extensive deconvolution analysis, was employed for the quantitative assessment of protein particles sizes, especially aimed at the quantification of 0.2-10 μm diameter subvisible particles (SVPs). The qLD accurately estimated concentration distributions for silica beads with diameters ranging from 0.2 to 10 μm that h  ...[more]

Similar Datasets

| S-EPMC7818018 | biostudies-literature
| S-EPMC7683618 | biostudies-literature
| S-EPMC10665318 | biostudies-literature
| S-EPMC9325253 | biostudies-literature
| S-EPMC4416045 | biostudies-literature
| S-EPMC4526581 | biostudies-literature
2015-04-09 | E-GEOD-60029 | biostudies-arrayexpress
| S-EPMC3904146 | biostudies-literature
| S-EPMC4711642 | biostudies-literature
2024-06-12 | GSE219069 | GEO