Intra-breath arterial oxygen oscillations detected by a fast oxygen sensor in an animal model of acute respiratory distress syndrome.


ABSTRACT:

Background

There is considerable interest in oxygen partial pressure (Po2) monitoring in physiology, and in tracking Po2 changes dynamically when it varies rapidly. For example, arterial Po2 ([Formula: see text]) can vary within the respiratory cycle in cyclical atelectasis (CA), where [Formula: see text] is thought to increase and decrease during inspiration and expiration, respectively. A sensor that detects these [Formula: see text] oscillations could become a useful diagnostic tool of CA during acute respiratory distress syndrome (ARDS).

Methods

We developed a fibreoptic Po2 sensor (<200 µm diameter), suitable for human use, that has a fast response time, and can measure Po2 continuously in blood. By altering the inspired fraction of oxygen ([Formula: see text]) from 21 to 100% in four healthy animal models, we determined the linearity of the sensor's signal over a wide range of [Formula: see text] values in vivo. We also hypothesized that the sensor could measure rapid intra-breath [Formula: see text] oscillations in a large animal model of ARDS.

Results

In the healthy animal models, [Formula: see text] responses to changes in [Formula: see text] were in agreement with conventional intermittent blood-gas analysis (n=39) for a wide range of [Formula: see text] values, from 10 to 73 kPa. In the animal lavage model of CA, the sensor detected [Formula: see text] oscillations, also at clinically relevant [Formula: see text] levels close to 9 kPa.

Conclusions

We conclude that these fibreoptic [Formula: see text] sensors have the potential to become a diagnostic tool for CA in ARDS.

SUBMITTER: Formenti F 

PROVIDER: S-EPMC4364062 | biostudies-literature | 2015 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Intra-breath arterial oxygen oscillations detected by a fast oxygen sensor in an animal model of acute respiratory distress syndrome.

Formenti F F   Chen R R   McPeak H H   Murison P J PJ   Matejovic M M   Hahn C E W CE   Farmery A D AD  

British journal of anaesthesia 20150128 4


<h4>Background</h4>There is considerable interest in oxygen partial pressure (Po2) monitoring in physiology, and in tracking Po2 changes dynamically when it varies rapidly. For example, arterial Po2 ([Formula: see text]) can vary within the respiratory cycle in cyclical atelectasis (CA), where [Formula: see text] is thought to increase and decrease during inspiration and expiration, respectively. A sensor that detects these [Formula: see text] oscillations could become a useful diagnostic tool o  ...[more]

Similar Datasets

| S-EPMC5587703 | biostudies-literature
| S-EPMC3941248 | biostudies-literature
| 2399547 | ecrin-mdr-crc
| S-EPMC4021554 | biostudies-other
| S-EPMC4654508 | biostudies-literature
| S-EPMC8280652 | biostudies-literature
| S-EPMC7815528 | biostudies-literature
| S-EPMC5851805 | biostudies-literature
| S-EPMC10960461 | biostudies-literature
| S-EPMC8746032 | biostudies-literature