Dosimetric verification by using the ArcCHECK system and 3DVH software for various target sizes.


ABSTRACT:

Objective

To investigate the usefulness of the 3DVH software with an ArcCHECK 3D diode array detector in newly designed plans with various target sizes.

Methods

The isocenter dose was measured with an ion-chamber and was compared with the planned and 3DVH predicted doses. The 2D gamma passing rates were evaluated at the diode level by using the ArcCHECK detector. The 3D gamma passing rates for specific regions of interest (ROIs) were also evaluated by using the 3DVH software. Several dose-volume histograms (DVH)-based predicted metrics for all structures were also obtained by using the 3DVH software.

Results

The isocenter dose deviation was <1% in all plans except in the case of a 1 cm target. Besides the gamma passing rate at the diode level, the 3D gamma passing rate for specific ROIs tended to decrease with increasing target size; this was more noticeable when a more stringent gamma criterion was applied. No correlation was found with the gamma passing rates and the DVH-based metrics especially in the ROI with high-dose gradients.

Conclusions

Delivery quality assurance by using 3DVH and ArcCHECK can provide substantial information through a simple and easy approach, although the accuracy of this system should be judged cautiously.

SUBMITTER: Song JH 

PROVIDER: S-EPMC4373958 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature

altmetric image

Publications

Dosimetric verification by using the ArcCHECK system and 3DVH software for various target sizes.

Song Jin Ho JH   Shin Hun-Joo HJ   Kay Chul Seung CS   Son Seok Hyun SH  

PloS one 20150325 3


<h4>Objective</h4>To investigate the usefulness of the 3DVH software with an ArcCHECK 3D diode array detector in newly designed plans with various target sizes.<h4>Methods</h4>The isocenter dose was measured with an ion-chamber and was compared with the planned and 3DVH predicted doses. The 2D gamma passing rates were evaluated at the diode level by using the ArcCHECK detector. The 3D gamma passing rates for specific regions of interest (ROIs) were also evaluated by using the 3DVH software. Seve  ...[more]

Similar Datasets

| S-EPMC10590446 | biostudies-literature
| S-EPMC10075462 | biostudies-literature
| S-EPMC2921183 | biostudies-literature
| S-EPMC5645924 | biostudies-literature
| S-EPMC5549624 | biostudies-other
| S-EPMC9180901 | biostudies-literature
| S-EPMC5874088 | biostudies-other
| S-EPMC5691007 | biostudies-other
| S-EPMC1144171 | biostudies-other
| S-EPMC6769018 | biostudies-literature