Multiplexed analysis of steroid hormones in human serum using novel microflow tile technology and LC-MS/MS.


ABSTRACT: A novel microfluidic chromatography device coupled with tandem mass spectrometry (LC-MS/MS) was utilized for the multiplex analysis of 5 steroids (testosterone, dihydrotestosterone, progesterone, cortisol, cortisone) in human serum. The use of microfluidics allowed for reduction of the chromatographic flow rate to 3?l/min with overall method run times comparable to standard flow LC-MS/MS methods reported in the literature, corresponding to a 150 fold decrease in solvent consumption. Furthermore, a simple sample preparation protocol was employed requiring injection of only 0.5?l of sample, corresponding to a 100-400 fold increase in on-column sensitivity as compared to published standard flow assays. The measured LOQ for both testosterone and progesterone was 0.4ng/mL, representing an improvement over reported literature values obtained by standard flow methods employing comparable sample preparation and large injection volumes. The LOQs for cortisol (1.9ng/mL), cortisone (0.3ng/mL), and dihydrotestosterone (1.4ng/mL) were all within a biologically relevant range. A comparison of clinical serum samples was performed for the analysis of testosterone using this microfluidic LC-MS/MS assay and the Beckman Access II automated antibody-based measurement system. The immunoassay results were systematically higher due to matrix interference which was easily resolved with the increased chromatographic resolution obtained in the microflow LC-MS/MS assay.

SUBMITTER: Broccardo CJ 

PROVIDER: S-EPMC4391816 | biostudies-literature | 2013 Sep

REPOSITORIES: biostudies-literature

altmetric image

Publications

Multiplexed analysis of steroid hormones in human serum using novel microflow tile technology and LC-MS/MS.

Broccardo Carolyn J CJ   Schauer Kevin L KL   Kohrt Wendy M WM   Schwartz Robert S RS   Murphy James P JP   Prenni Jessica E JE  

Journal of chromatography. B, Analytical technologies in the biomedical and life sciences 20130704


A novel microfluidic chromatography device coupled with tandem mass spectrometry (LC-MS/MS) was utilized for the multiplex analysis of 5 steroids (testosterone, dihydrotestosterone, progesterone, cortisol, cortisone) in human serum. The use of microfluidics allowed for reduction of the chromatographic flow rate to 3μl/min with overall method run times comparable to standard flow LC-MS/MS methods reported in the literature, corresponding to a 150 fold decrease in solvent consumption. Furthermore,  ...[more]

Similar Datasets

| S-EPMC11322771 | biostudies-literature
| S-EPMC7447131 | biostudies-literature
| S-EPMC10836901 | biostudies-literature
| S-EPMC9921923 | biostudies-literature
| S-EPMC7112139 | biostudies-literature
| S-EPMC8772585 | biostudies-literature
| S-EPMC5636693 | biostudies-literature
| S-EPMC6069520 | biostudies-literature
| S-HECA171 | biostudies-other
| S-EPMC9314307 | biostudies-literature